MOSAIQ recallZ-2406-2026 · Elekta, Inc. · initiated May 21, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.

FDA root cause category: Software design

What the company is doing

On May 21, 2026, Elekta issued a "Urgent Medical Device Notification to affected to consignees via E-Mail. Elekta asked consignees to take the following actions: 1. Please review this notification and identify if you are running the effected software version listed above. 2. If an Application Communication Time Out error message appears during a BLS plan transfer, please refer to the attached appendix. 3. Please acknowledge receipt of this notification via the Elekta Care Community or complete this form and return it to Elekta immediately upon receipt, but no later than within 30 days. 4. Please post this notice in a place accessible to all users, e.g. Instructions for Use, until this action is closed. 5. Advise the appropriate personnel working with this product of the content of this letter. 6. If affected product has been transferred to a different location, please ensure that location and personnel receive a copy of this notification.

Recall details

Recall number
Z-2406-2026
Status
Open, Classified
Initiated by the company
May 21, 2026
Posted by the FDA
Jun 11, 2026
Recalling firm
Elekta, Inc.
Quantity
12 programs
Product code
IYE
FDA submissions
K223229
Distribution
Worldwide - US Nationwide distribution in the states of WI, FL, LA, NY, NY, NJ, KS, FL, IA, NM, TX and the countries of Bahrain, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, India, Israel, Italy, Japan, Lithuania, Malta, Mexico, Netherlands, Pakistan, Panama, Poland, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom.

Product description

MOSAIQ Oncology Information System

Code information (models, lots, serial numbers)

Software Version: MOSAIQ 3.2/UDI: 07340201500071