Skip to content

MOSAIQRecalls & reported problems across all models

Elekta Solutions AB · 6 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

6 open recalls name at least one MOSAIQ model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
185
Deaths reported
0
Injuries reported
4
Recalls
7

We found 6 models of MOSAIQ by Elekta Solutions AB in the FDA's device registry (GUDID). Together they are named in 185 adverse event reports in the FDA's MAUDE database, received between Mar 6, 2020 and Sep 14, 2026 — classified as 4 injuries, 179 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (185 reports).

7 FDA recall records name at least one of these models by its device identifier, 6 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 5 reportsDec 2023: 5 reportsJan 2024: 3 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 5 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 2 reportsJan 2025: 6 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 7 reportsJun 2025: 16 reportsJul 2025: 2 reportsAug 2025: 3 reportsSep 2025: 3 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

MOSAIQ recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. May 21, 2026Open, Classified

    Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.

    Z-2406-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Dec 10, 2025Open, Classified

    Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

    Z-1078-2026 · Root cause: Software change control · names 2 of these models · Recall details · FDA record ↗

  3. Sep 17, 2025Open, Classified

    When appending a care plan that contains one or more wave medication orders, the occurrence and frequency of the appended orders may not match the intended schedule.

    Z-0503-2026 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  4. Jan 15, 2025Open, Classified

    Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.

    Z-1162-2025 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  5. Sep 2, 2021Open, Classified

    A drug strength in MOSAIQ can be changed during the ordering process and approved, but the saved drug strength reverts to the original value.

    Z-2591-2021 · Root cause: Software design · Recall details · FDA record ↗

  6. Apr 20, 2021Terminated

    Malware attack

    Z-1783-2021 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗

  7. Sep 24, 2019Open, Classified

    The user may inadvertently enter Metric values into Height and Weight fields labeled with US Standard Units.

    Z-1597-2020 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

Most reported MOSAIQ problems

Device problems

  • Computer Software Problem185
  • Communication or Transmission Problem3
  • Use of Device Problem3
  • Intermittent Communication Failure2
  • Application Network Problem1
  • Application Program Problem: Medication Error1
  • Device Handling Problem1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Insufficient Information167
  • No Clinical Signs, Symptoms or Conditions139
  • No Information7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1
  • Unintended Radiation Exposure1

MOSAIQ models

6 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent MOSAIQ adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Elekta Solutions AB products

All Elekta Solutions AB devices →

Sources and updates