Skip to content

ARTIS pheno, Model 10849000. interventional fluoroscopic… recallZ-2205-2026 · Siemens Medical Solutions USA, Inc · initiated Mar 9, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Distribution of Customer Advisory Notice (CAN) to all affected customers began by Siemens on March 9, 2026.

Recall details

Recall number
Z-2205-2026
Status
Open, Classified
Initiated by the company
Mar 9, 2026
Posted by the FDA
Jun 4, 2026
Recalling firm
Siemens Medical Solutions USA, Inc
Product code
OWB
FDA submissions
K241572
Distribution
US and Worldwide

Product description

ARTIS pheno, Model 10849000. interventional fluoroscopic x-ray system

Code information (models, lots, serial numbers)

UDI: 4056869046877

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K241572 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 98992).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page