Skip to content

Impella Heart Pump recallZ-2601-2026 · Abiomed, Inc. · initiated May 22, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abiomed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for thrombus formation during prolonged use of the introducer.

FDA root cause category: Under Investigation by firm

What the company is doing

Abiomed notified consignees on about 05/22/2026, via letter titled "URGENT MEDICAL DEVICE RECALL (REMOVAL)." Consignees were made aware of the issue and that if thrombus formation were to occur, the user may encounter difficulty aspirating the sheath sidearm or valve and may observe thrombus in the syringe or sidearm post aspiration or around the valve post CP pump removal. They were instructed to review inventory for affected units and quarantine them, arrange for their return using the provided shipment label from Sedgwick, and complete and return the provided business response form. Consignees were also instructed to forward the notification to anyone in the facility that needs to be informed, if any of the subject products have been forwarded to another facility, contact that facility and provide them with the notification, and to post a copy of the notification in a visible area for awareness.

Recall details

Recall number
Z-2601-2026
Status
Open, Classified
Initiated by the company
May 22, 2026
Posted by the FDA
Jun 25, 2026
Recalling firm
Abiomed, Inc.
Quantity
168 units
Product code
DYB
FDA submissions
K241708
Distribution
Worldwide - US Nationwide distribution in the states of FL, IL, MN, NJ, NY, VA and the country of Japan.

Product description

Abiomed Impella CP Set with SmartAssist (10th Generation) containing affected 14Fr Low Profile Introducer Kits. Product Codes: 1000413, 1000834.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

1. Impella Set Product Code: 1000413; GTIN: 00813502013566; Serial numbers: 649727, 649728, 649729, 649730, 649731, 654349, 654348, 654342, 654341, 654340, 654339, 654338, 654336, 654335, 654334, 654333, 653010, 653009, 653008, 653007, 653006, 653005, 653004, 653003, 653002, 653001, 655735, 655733, 654357, 654356, 654382, 654383, 654381, 654380, 654378, 654367, 654366, 654364, 654365, 654362, 654363, 654361, 654360, 654359, 654358, 654379, 655734, 654386, 654355, 654385, 654384, 654353, 654352, 654351, 654350, 654387, 657649, 657648, 657646, 657647, 657645, 657644, 655742, 655739, 655741, 655740, 655738, 655737, 655736, 656743, 656738, 653050, 653049, 653048, 653047, 653045, 653044, 653043, 653042, 653041. Introducer Product Code: 2000684; Batch Numbers: 2026724558, 2027818724. 2. Product Code: 1000834. GTIN: 00813502013948. Serial numbers: 653474, 653471, 653485, 653470, 653469, 653468, 653467, 653465, 653466, 653463, 653464, 653482, 653483, 653484, 653481, 653490, 653462, 653478, 653479, 653489, 653488, 653487, 653486, 653476, 653477, 656002, 656001, 656000, 655998, 655997, 655993, 655996, 655995, 655994, 656737, 656736, 656735, 656734, 656733, 656730, 656731, 656729, 656728, 656346, 656343, 656342, 656341, 656340, 656338, 659804, 659795, 656347, 650264, 659801, 659802, 659800, 659799, 659798, 656363, 656367, 656362, 656361, 656360, 656359, 656358, 659794, 659793, 659791, 659792, 659790, 659789, 659788, 659786, 659787, 659785, 661408, 661407, 661405, 661404, 661392, 661391, 661389, 661403, 661402, 661401, 661400, 662657, 662654. Introducer Product Code: 2000785; Batch Numbers: 2026787668, 2027818660. Distributed in Japan.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 99002).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page