Impella Heart PumpRecalls & reported problems across all models

Abiomed, Inc. · 34 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

42 open recalls name at least one Impella Heart Pump model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
30,421
Deaths reported
3,817
Injuries reported
15,140
Recalls
46

We found 34 models of Impella Heart Pump by Abiomed, Inc. in the FDA's device registry (GUDID). Together they are named in 30,421 adverse event reports in the FDA's MAUDE database, received between Nov 23, 2015 and Sep 30, 2026 — classified as 3,817 deaths, 15,140 injuries, 11,464 malfunctions.

The most frequently reported specific device problem is “Restricted Flow rate” (2,416 reports).

46 FDA recall records name at least one of these models by its device identifier, 42 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 384 reportsDec 2023: 296 reportsJan 2024: 609 reportsFeb 2024: 70 reportsMar 2024: 1,535 reportsApr 2024: 1,214 reportsMay 2024: 924 reportsJun 2024: 716 reportsJul 2024: 438 reportsAug 2024: 2,849 reportsSep 2024: 1,079 reportsOct 2024: 2,043 reportsNov 2024: 2,017 reportsDec 2024: 384 reportsJan 2025: 544 reportsFeb 2025: 336 reportsMar 2025: 538 reportsApr 2025: 416 reportsMay 2025: 751 reportsJun 2025: 857 reportsJul 2025: 817 reportsAug 2025: 428 reportsSep 2025: 819 reportsOct 2025: 465 reportsNov 2025: 570 reportsDec 2025: 640 reportsJan 2026: 961 reportsFeb 2026: 1,551 reportsMar 2026: 1,357 reportsApr 2026: 648 reportsMay 2026: 386 reportsJun 2026: 708 reportsJul 2026: 179 reportsAug 2026: 216 reportsSep 2026: 319 reportsOct 2026: 0 reports2024202520262,849
Last 36 months. The most recent month may be incomplete.

Impella Heart Pump recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Aug 12, 2026Open, Classified

    The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

    Z-3150-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Aug 12, 2026Open, Classified

    The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

    Z-3149-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Aug 12, 2026Open, Classified

    The potential for purge cassette recognition issues due to the failure of the purge flag component of the purge pressure sensor assembly.

    Z-3148-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  4. Jun 11, 2026Open, Classified

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

    Z-2686-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  5. Jun 11, 2026Open, Classified

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

    Z-2685-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  6. Jun 11, 2026Open, Classified

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

    Z-2684-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  7. Jun 11, 2026Open, Classified

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

    Z-2683-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Jun 11, 2026Open, Classified

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

    Z-2682-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  9. Jun 11, 2026Open, Classified

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

    Z-2681-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  10. May 22, 2026Open, Classified

    Potential for thrombus formation during prolonged use of the introducer.

    Z-2601-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  11. May 22, 2026Open, Classified

    Potential for thrombus formation during prolonged use of the introducer.

    Z-2599-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  12. May 18, 2026Open, Classified

    Exposure to the low purge pressure occurrence may result in persistent low purge pressure alarms and, in some cases, interruption or loss of mechanical circulatory support.

    Z-2457-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  13. May 14, 2026Open, Classified

    Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.

    Z-2360-2026 · Root cause: Under Investigation by firm · names 4 of these models · Recall details · FDA record ↗

  14. Apr 20, 2026Open, Classified

    A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.

    Z-2211-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  15. Feb 18, 2026Open, Classified

    Increased risk of purge leaks with Generation 1 purge cassettes.

    Z-1589-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  16. Feb 16, 2026Open, Classified

    Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.

    Z-1591-2026 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  17. Jan 27, 2026Open, Classified

    Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.

    Z-1470-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  18. Jan 27, 2026Open, Classified

    Differential pressure (dP) sensor in Impella RP devices may malfunction, causing sensor values to drift.

    Z-1469-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  19. Dec 4, 2025Open, Classified

    Device packaged in incorrect outer box carton.

    Z-1030-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  20. Oct 1, 2025Open, Classified

    Potential cybersecurity vulnerabilities related to the operating system in the Automated Impella Controller (AIC).

    Z-0069-2026 · Root cause: Software design · names 4 of these models · Recall details · FDA record ↗

  21. Sep 16, 2025Open, Classified

    Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.

    Z-0136-2026 · Root cause: Device Design · names 4 of these models · Recall details · FDA record ↗

  22. Aug 20, 2025Open, Classified

    Specific Automated Impella Controllers (AICs) that have a Pump Driver Circuit Assembly do not meet current specifications. This may lead to decreased pump performance or pump stop and trigger an Impella Failure or Impella Stopped Controller Failure alarm.

    Z-2568-2025 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  23. Jun 23, 2025Open, Classified

    A potential issue with the Automated Impella Controller (AIC) not detecting an Impella pump when it is connected

    Z-2164-2025 · Root cause: Device Design · names 4 of these models · Recall details · FDA record ↗

  24. Dec 12, 2024Open, Classified

    Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.

    Z-0830-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  25. Dec 12, 2024Open, Classified

    Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.

    Z-0829-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  26. Aug 5, 2024Open, Classified

    A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

    Z-2965-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  27. Aug 5, 2024Open, Classified

    A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

    Z-2964-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  28. Aug 5, 2024Open, Classified

    A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

    Z-2963-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  29. Aug 5, 2024Open, Classified

    A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

    Z-2959-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  30. May 31, 2024Open, Classified

    Nine (9) Impella CP pumps failed inspection and were inadvertently released.

    Z-2319-2024 · Root cause: Release of Material/Component prior to receiving test results · Recall details · FDA record ↗

…and 16 older recalls, listed on each model's page.

Most reported Impella Heart Pump problems

Device problems

  • Adverse Event Without Identified Device or Use Problem13,234
  • Restricted Flow rate2,416
  • Device Sensing Problem2,070
  • Patient-Device Incompatibility1,676
  • Patient Device Interaction Problem1,392
  • Failure to Advance1,229
  • Positioning Failure1,169
  • Malposition of Device1,137

Patient problems

  • No Clinical Signs, Symptoms or Conditions12,192
  • Hemorrhage/Blood Loss/Bleeding6,635
  • Ischemia2,957
  • Hemolysis2,645
  • Hematoma2,036
  • Thrombosis/Thrombus1,686
  • Tachycardia1,280
  • Arrhythmia955

Impella Heart Pump models

34 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Impella Heart Pump adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.