Skip to content

AMD-RITMED recallZ-2459-2026 · AMD Medicom Inc. · initiated May 8, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from AMD Medicom Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.

FDA root cause category: Material/Component Contamination

What the company is doing

On 5/8/2026, recall notices were emailed to customers who were asked to do the following: 1) Discontinue use, quarantine and destroy affected devices. 2) Notify any customers or facilities to whom you have distributed this product. 3) Complete and return the response form via email to firm's customer service at [email protected] On 6/5/2026, recall notices with all affected lots were emailed to customers.

Recall details

Recall number
Z-2459-2026
Status
Open, Classified
Initiated by the company
May 8, 2026
Posted by the FDA
Jun 16, 2026
Recalling firm
AMD Medicom Inc.
Quantity
10
Product code
EFQ
FDA submissions
K830269
Distribution
US: MT, NC, VA, CA, TX, OH, TN, IL, RI

Product description

Ritmed Neurological Sponges, 0.75x0.75in, Sterile, REF: 85995C

Code information (models, lots, serial numbers)

UDI-DI: 10686864044465, Lot: 17936

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 99006).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page