Breathing circuit set, coaxial recallZ-2463-2026 · Hamilton Medical AG · initiated May 29, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Hamilton Medical AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement

FDA root cause category: Component design/selection

What the company is doing

On 5/29/2026, correction notices were emailed to customers who were asked to do the following: 1) Inform all potential users of affected devices within your facility about this issue. 2) Inform all affected end users and/or sub-distributors and provide them with this notice. 5) Post this notice to ensure all personnel are informed. 6) Complete and return Facility Reply Form via email to [email protected] Customers with questions can contact the firm at the email above. Initially customers were informed to continue use and perform actions to mitigate risk, but later the updated notice called for affected device removal. On 7/8/2026, updated recall notices were emailed and mailed to healthcare facilities, physicians and end users who were informed of the following: 1) Discontinue use and quarantine affected devices. 2) Do not follow previous instructions for product modification. 3) Inform all affected end users and/or sub-distributors and provide them with this notice. 4) Contact firm if you have affected product in inventory. - Discard any affected product(s) and confirm destruction. Or - Contact Hamilton Medical for a return material authorization number and return label. -Contact your sales representative to acquire replacement product. 5) Complete and return the Facility Reply Form.

Recall details

Recall number
Z-2463-2026
Status
Open, Classified
Initiated by the company
May 29, 2026
Posted by the FDA
Jun 30, 2026
Recalling firm
Hamilton Medical AG
Quantity
27,270 pcs
Product code
BZO
Distribution
US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.

Product description

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Code information (models, lots, serial numbers)

REF/UDI-DI: 260127/07630002802956, 17630002802953; 260128/07630002802963, 17630002802960; 260167/07630002802970, 17630002802977; 260168/07630002802987, 17630002802984. Lot numbers between 200379 and 205050 (including both.)