Comprehensive recallZ-2529-2026 · Zimmer, Inc. · initiated May 20, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE RECALL CORRECTION notification letter dated 5/26/26 was sent to customers. Distributor Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement for each return and send to [email protected]. 3. Return the Additional Accounts form to [email protected]. a. Review the list of hospitals and/or surgeons included with the email notification sent to your facility, which includes a list of hospitals and/or surgeons that have already been notified of this recall. b. Identify whether there are any additional hospitals and/or surgeons who are current users of the affected products that Zimmer Biomet has not notified and list these accounts on the Additional Accounts form. Please provide the form in Excel format. c. If there are no additional accounts or surgeons to notify, please indicate that there are no additional accounts, or indicate None or NA on the form. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 a.m. and 5:00 p.m. Eastern, Monday through Friday. Alternatively, your questions may be emailed to [email protected]. Risk Manager Responsibilities: 1. Review this correction notice and ensure that affected personnel are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 3. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 4. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 am and 5:00pm EST, Monday through Friday. Alternatively, y
Recall details
- Recall number
- Z-2529-2026
- Status
- Open, Classified
- Initiated by the company
- May 20, 2026
- Posted by the FDA
- Jun 23, 2026
- Recalling firm
- Zimmer, Inc.
- Quantity
- 429 units
- Product code
- PHX
- FDA submissions
- K241043
- Distribution
- US Nationwide distribution.
Product description
Off-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040242
Code information (models, lots, serial numbers)
Lot Code: Model No. 110040242 ; UDI-DI (01)00889024681989(17)350601(10)099471 ; Lot Number 099471 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350614(10)099473 ; Lot Number 099473 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350804(10)099474 ; Lot Number 099474 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350806(10)099475 ; Lot Number 099475
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Comprehensive 110040242Zimmer, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented1 recall0 reports
Product lines
Other devices under the same FDA submission
5 devices cleared under K241043 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Comprehensive 110040240Zimmer, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented1 recall2 reports
- Alliance 110040444Zimmer, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented1 recall1 report
- Alliance 110040476Zimmer, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented1 recall1 report
- Comprehensive 110040241Zimmer, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented1 recall0 reports
- Comprehensive 110040435Zimmer, Inc. · Prosthesis, Shoulder, Non-Constrained, Metal/Polymer Cemented1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 99045).
- Z-2527-2026 · Open, ClassifiedOff-Axis Comprehensive Shoulder System Small, Augment, Off-Axis, Reamer Guide, Model/Catalog Number: 110040240
- Z-2528-2026 · Open, ClassifiedOff-Axis Comprehensive Shoulder System, Medium, Augment, Off-Axis, Reamer Guide Model/Catalog Number: 110040241
- Z-2530-2026 · Open, ClassifiedOff-Axis Alliance Glenoid, Left, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040444
- Z-2531-2026 · Open, ClassifiedOff-Axis Alliance Glenoid, Right, 4-Peg, Augment, Reamer Guide, Model/Catalog Number: 110040476
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page