5008X Blood Tubing Set, 24 units per case. Includes the… recallZ-2702-2026 · Fresenius Medical Care Holdings, Inc. · initiated May 11, 2026
Class II Open, Classified All recalls posted in July 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the 5008X CAREsystem.
FDA root cause category: Process control
What the company is doing
Fresenius Medical Care notified consignees on about 05/11/2026 via "IMPORTANT MEDICAL DEVICE INFORMATION" letter. Consignees were informed of the issue of potential Online Port Errors and additional information to support proper connection of the SafeLine of the 5008X CAREset to the substitution fluid port of the 5008X CAREsystem. When pushing the substitution fluid connector (SafeLineTM) into the substitution fluid port, firmly insert the substitution fluid connector until you feel resistance. Check that there is no visible gap between the ring of the connector and the module. In the case of a visible gap between the ring of the connector and the module surface, apply pressure and push the connector to close the gap before automatic closure of the substitution fluid port door. Fresenius Medical Care is assessing improvements to restore the usual handling and user experience. FMCRTG is not requesting return of any product. ***Update*** On 06/292026, the firm sent a follow-up letter to consignees via First Class Mail. The updated letter states the firm has initiated shipments of limited quantities of new unaffected product. Customers are instructed to preferentially use unaffected lots of the tubing sets. If the customer needs to administer treatment and there are no alternatives available, they may use the affected lots of tubing sets to provide therapy, but only if they follow the specific instructions for using these affected sets. Unaffected lots of product do not require additional effort to seat the connector into the port. ***Update*** On 08/10/2026, the firm sent a follow up letter to consignees notifying them that product stock is available to replace affected products and the notification included specific lot numbers affected.
Recall details
- Recall number
- Z-2702-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- May 11, 2026
- Posted by the FDA
- Jul 8, 2026
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Quantity
- 113,544 units (4,731 cases)
- Product code
- KDI
- FDA submissions
- K243505
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, GA, IL, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NM, NY, OH, PA, TX, VA.
Product description
5008X Blood Tubing Set, 24 units per case. Includes the below model numbers and descriptions. 1. 5008X HD Pre-Flush Standard Blood Tubing Set; Model Number: 03-5110-6. 2. 5008X HD/HDF Standard Blood Tubing Set; Model Number: 03-5300-3. 3. 5008X HD/HDF With CLiC Blood Tubing Set; Model Number: 03-5300-3C. 4. 5008X HD/HDF With Twister Blood Tubing Set; Model Number: 03-5350-8.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Sets manufactured after February 26, 2026 1. Model Number: 03-5110-6; UDI-DI: 00840861102556; Lot Number: 26BR01177. 2. Model Number: 03-5300-3; UDI-DI: 00840861102518; Lot Numbers: 26BR01184, 26BR01202, 26CR01025, 26CR01039, 26CR01803, 26BR01185, 26BR01214, 26CR01026, 26CR01040, 26CR01804, 26BR01188, 26BR01215, 26CR01027, 26CR01041, 26DR01134, 26BR01189, 26BR01805, 26CR01029, 26CR01043, 26BR01201, 26CR01003, 26CR01034, 26CR01800. 3. Model Number: 03-5300-3C; UDI-DI: 00840861102532; Lot Numbers: 26CR01018, 26CR01020, 26DR01129, 26DR01150, 26CR01019, 26CR01021, 26DR01130, 26DR01151. 4. Model Number: 03-5350-8; UDI-DI: 00840861102525; Lot Numbers: 26CR01017, 26DR01077, 26DR01079, 26DR01080.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- 5008X 03-5300-3Fresenius Medical Care Renal Therapies Group, LLC · Set, Tubing, Blood, With And Without Anti-Regurgitation Valve2 recalls23 reports
- 5008X 03-5300-3CFresenius Medical Care Renal Therapies Group, LLC · Set, Tubing, Blood, With And Without Anti-Regurgitation Valve2 recalls6 reports
- 5008X 03-5350-8Fresenius Medical Care Renal Therapies Group, LLC · Set, Tubing, Blood, With And Without Anti-Regurgitation Valve2 recalls0 reports
- 5008X 03-5110-6Fresenius Medical Care Renal Therapies Group, LLC · Set, Tubing, Blood, With And Without Anti-Regurgitation Valve2 recalls0 reports
Other devices under the same FDA submission
1 device cleared under K243505 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- 5008X 04-3000-9Fresenius Medical Care Renal Therapies Group, LLC · Set, Tubing, Blood, With And Without Anti-Regurgitation Valve1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.