Skip to content

5008X 03-5350-8Recalls & reported problems

Fresenius Medical Care Renal Therapies Group, LLC · 5008X HD/HDF - with Twister Blood Tubing Set · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

4 other recalls cover devices cleared under the same FDA submission (K243505); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Aug 10, 2026Open, Classified

    Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.

    Z-3087-2026 · Root cause: Component change control · FDA record ↗

  2. May 11, 2026Open, Classified

    Potential for Online Port Errors due to connection issues when connecting the SafeLine of the 5008X CAREset to the substitution fluid port of the 5008X CAREsystem.

    Z-2702-2026 · Root cause: Process control · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K243505. One submission often covers many models, so check the product and lot details in each record.

  • Z-3086-2026 · Aug 10, 2026 · Open, Classified
    5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C
  • Z-3085-2026 · Aug 10, 2026 · Open, Classified
    5008X HD/HDF Standard Blood Tubing Set Part Number 03-5300-3
  • Z-3084-2026 · Aug 10, 2026 · Open, Classified
    5008X Standard HD Pre-Flush Standard Blood Tubing Set Part Number 03-5110-6
  • Z-0936-2026 · Oct 25, 2025 · Open, Classified
    5008X CAREsystem +CLiC +CDX;

Other Set, Tubing, Blood, With And Without Anti-Regurgitation Valve devices

All Set, Tubing, Blood, With And Without Anti-Regurgitation Valve devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.