ARTIS icono biplane. Model Number: 11327600 recallZ-2641-2026 · Siemens Medical Solutions USA, Inc · initiated May 20, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits.
FDA root cause category: Software design
What the company is doing
Siemens notified consignees on about 05/19/2026 via letter. Consignees were instructed to ensure all users of the affected products within each organization and others who may need to be informed will receive the relevant safety information provided within the notice and comply with the recommendations therein, maintain the recall notification with product related records and until the corrective measures have been completed. Siemens will be releasing a software update to correct the issue via an Update Instruction AX018/26/S. All affected systems will be corrected by Siemens Healthineers. The service organization will contact consignees in the near future to arrange a date to perform the corrective action. Additionally, consignees were instructed to confirm and acknowledge receipt of the recall notification digitally.
Recall details
- Recall number
- Z-2641-2026
- Status
- Open, Classified
- Initiated by the company
- May 20, 2026
- Posted by the FDA
- Jul 1, 2026
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 85 units
- Product code
- OWB
- FDA submissions
- K241572
- Distribution
- US Nationwide distribution.
Product description
ARTIS icono biplane. Model Number: 11327600.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model Number: 11327600. UDI-DI: 4056869063317. Serial Numbers: 184246, 184241, 184086, 184178, 184108, 184099, 184070, 184127, 184095, 184167, 184184, 184205, 184199, 184221, 184126, 184047, 184093, 184044, 184096, 184214, 184100, 184119, 184073, 184188, 184134, 184150, 184081, 184153, 184085, 184243, 184117, 184045, 184159, 184151, 184177, 184079, 184087, 184169, 184116, 184157, 184163, 184123, 184090, 184118, 184140, 184230, 184224, 184195, 184248, 184225, 184190, 184106, 184046, 184078, 184200, 184142, 184138, 184166, 184152, 184113, 184186, 184196, 184105, 184102, 184147, 184120, 184121, 184124, 184211, 184233, 184216, 184050, 184067, 184112, 184122, 184220, 184089, 184097, 184234, 184103, 184098, 184084, 184183, 184179, 184109
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- ARTIS icono biplane 11327600Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls2 deaths55 reports
Other devices under the same FDA submission
3 devices cleared under K241572 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ARTIS icono floor 11327700Siemens Healthcare GmbH · System, x-ray, tomography, computed15 recalls24 reports
- ARTIS pheno 10849000Siemens Healthcare GmbH · Interventional fluoroscopic x-ray system14 recalls64 reports
- ARTIS icono ceiling 11328100Siemens Healthcare GmbH · System, x-ray, angiographic9 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 99079).
- Z-2642-2026 · Open, ClassifiedARTIS icono ceiling. Model Number: 11328100.
- Z-2643-2026 · Open, ClassifiedARTIS icono floor. Model Number: 11327700.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page