Skip to content

Cyberknife Treatment Deliver System equipped with the… recallZ-2644-2026 · Accuray Incorporated · initiated Jun 16, 2026

Class II Open, Classified All recalls posted in July 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Accuray Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

FDA root cause category: Software design

What the company is doing

Accuray issued an Urgent Field Safety Notice to its consignees via hand delivered letter on 6/16/2026. The notice explained the issue, risk to health, and provided the following safety instructions: "This condition may occur during collimator housing drop off or pick up while the collimator housing is positioned over the Xchange table. The system will interlock, stop motion, and display a message at the operator console to contact service. If an interlock occurs: (1) remove the patient from the treatment room before troubleshooting, (2) do not initiate treatment robot motion from the in-room control pendant, and (3) contact Accuray Service to safely release the collimator housing." Accuray is correcting the units. For questions, please contact Accuray Customer Support by phone or email, using the Service Request form available at http://www.accuray.com/service-requests.

Recall details

Recall number
Z-2644-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jun 16, 2026
Posted by the FDA
Jul 1, 2026
Recalling firm
Accuray Incorporated
Quantity
292 units
Product code
IYE
FDA submissions
K170788
Distribution
Worldwide - US Nationwide and and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kenya, Korea, Republic of, Liechtenstein, Lithuania, Luxembourg, Macao, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.

Product description

Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator

Code information (models, lots, serial numbers)

ALL Cyberknife systems running software version 11.X and higher that utilize the Exchange Table.

Other devices under the same FDA submission

1 device cleared under K170788 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.