CyberKnife® Treatment Delivery System 0660000Recalls & reported problems

Accuray Incorporated · The CyberKnife® Treatment Delivery System (CKTDS) is a computer-controlled radiation treatment delivery system for performing minimally invasive stereotactic radiosurgery and precision radiotherapy. The CyberKnife Treatment Delivery System is intended to work with Accuray Precision® Treatment Planning System and iDMS™ Data Management System. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00811376030016), received between Oct 28, 2021 and Aug 21, 2025 — classified as 3 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K170788); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Detachment of Device or Device Component2
  • Unintended System Motion2
  • Electrical /Electronic Property Problem1
  • Therapy Delivered to Incorrect Body Area1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Insufficient Information2
  • Nerve Damage1
  • Neuropathy1
  • Ulcer1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RECTAL TOXICITY DUE TO PROSTATE TREATMENT.

    Patient problems: Ulcer

    MAUDE report 22865305 ↗
  • InjuryJul 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    RECTAL TOXICITY DUE TO PROSTATE TREATMENT.

    Patient problems: Insufficient Information

    MAUDE report 22671925 ↗
  • MalfunctionOct 17, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MLC (MULTI-LEAF COLLIMATOR) DROPPED OFF OF THE SYSTEM AT PERCH AFTER AN ERROR OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20470488 ↗
  • MalfunctionJun 18, 2024Unintended System Motion; Unintended System Motion

    STANDARD TREATMENT COUCH ROLL JOINT WAS LOOSE AND CIRCLIP CAME OUT CAUSING THE COUCH TO ROTATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19562294 ↗
  • MalfunctionSep 8, 2023Unintended System Motion; Unintended System Motion

    THE STANDARD TREATMENT COUCH SNAP RING DROPPED, CAUSING THE COUCH TO ROLL DURING PATIENT ALIGNMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17710153 ↗
  • MalfunctionMar 21, 2023Unstable; Unstable

    THE CYBERKNIFE IMAGING HAD SHOWN A COUCH SAG MORE THAN 10 DEGREES FOR A PROSTATE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16585590 ↗
  • InjuryNov 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    SIDE EFFECT - ANESTHESIA DELAROSA (AD).

    Patient problems: Nerve Damage; Neuropathy

    MAUDE report 15773666 ↗
  • MalfunctionJul 22, 2022Therapy Delivered to Incorrect Body Area; Detachment of Device or Device Component; Therapy Delivered to Incorrect Body Area; Detachment of Device or Device Component

    CUSTOMER HAD A FAILED AQA (AUTOMATIC QUALITY ASSURANCE). FSE (FIELD SERVICE ENGINEER) FOUND THE FCA (FIXED COLLIMATOR ASSEMBLY) HAD DROPPED INSIDE THE COVERS.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15089280 ↗
  • MalfunctionOct 28, 2021Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT IS REPORTED THAT A TRANSFORMER WITHIN THE X-RAY GENERATOR OF CYBERKNIFE TREATMENT DELIVERY SYSTEM CAUGHT FIRE DURING INSTALLATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12711568 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K170788. One submission often covers many models, so check the product and lot details in each record.

  • Z-2644-2026 · Jun 16, 2026 · Open, Classified
    Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
  • Z-2640-2026 · Jun 16, 2026 · Open, Classified
    Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
  • Z-1023-2025 · Dec 23, 2024 · Completed
    CyberKnife Treatment Delivery System, REF 0660000 The CyberKnife System may be labeled as CyberKnife M6TM or VSI TM Robotic Surgery System. The CyberKnife Treatment Delivery Syste
  • Z-1272-2021 · Feb 26, 2021 · Terminated
    CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in

Other Accelerator, Linear, Medical devices

All Accelerator, Linear, Medical devices →