Introducer Kit for Impella recallZ-2679-2026 · Oscor Inc. · initiated Jun 10, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Oscor Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
FDA root cause category: Device Design
What the company is doing
On June 10, 2026, Oscor (an Integer subsidiary ) issued a Urgent Medical Device Recall notification to affected consignees. Oscor asked consignees to take the following actions: 1. Product removal is not required, and hospitals may continue to use existing inventory. 2. Physicians should continue to use appropriate clinical awareness, monitoring, and patient support as needed. 3. Consider a device exchange or using the repositioning sheath to minimize blood loss if potential leakage from the sidearm and under the sheath cap and along the hub score lines in the hub region is observed. 4.Review the information within this letter, complete all fields, sign, and return the attached business response form (BRF) (refer to Attachment 1) to [email protected]. 5. Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). 6.If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. 7. Post a copy of this notice in a visible area for awareness.
Recall details
- Recall number
- Z-2679-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 10, 2026
- Posted by the FDA
- Jul 10, 2026
- Recalling firm
- Oscor Inc.
- Quantity
- 1332 units
- Product code
- DYB
- FDA submissions
- K122084
- Distribution
- Distribution in the US state of MA
Product description
Abiomed REF: 0052-3021, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F, 23F. heart pump and accessories.
Code information (models, lots, serial numbers)
All Lots within Expiry/GTIN:00885672009786/00813502010558
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Introducer Kit for Impella® 0052-3021Oscor Inc. · INTRODUCER, CATHETER2 recalls17 reports
Product lines
Other devices under the same FDA submission
29 devices cleared under K122084 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Introducer Kit for Impella® 0052-3006Oscor Inc. · pacemaker lead Adaptor4 recalls223 deaths1,184 reports
- Introducer Kit for Impella® 0052-3006Oscor Inc. · INTRODUCER, CATHETER2 recalls4 deaths33 reports
- Introducer Kit for Impella® 0052-3025Oscor Inc. · INTRODUCER, CATHETER1 recall1 death32 reports
- Introducer Kit for Impella® 0052-3025Oscor Inc. · INTRODUCER, CATHETER1 recall10 reports
- SafeSheath® II SS6Oscor Inc. · INTRODUCER, CATHETER48 reports
- SafeSheath® II SS7Oscor Inc. · INTRODUCER, CATHETER42 reports
- SafeSheath® II SS9Oscor Inc. · INTRODUCER, CATHETER1 death21 reports
- Introducer Kit for Impella® 0052-3000Oscor Inc. · pacemaker lead Adaptor2 deaths19 reports
- Introducer Kit for Impella® 0052-3021Oscor Inc. · INTRODUCER, CATHETER1 death18 reports
- SafeSheath® II 667460-300Oscor Inc. · INTRODUCER, CATHETER10 reports
- Introducer Kit for Impella® 0052-3001Oscor Inc. · pacemaker lead Adaptor2 deaths6 reports
- SafeSheath® II SS8Oscor Inc. · INTRODUCER, CATHETER5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99102).
- Z-2678-2026 · Open, ClassifiedAbiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.
- Z-2680-2026 · Open, ClassifiedAbiomed REF: 0052-3006, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 23F. heart pump and accessories.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page