Skip to content

Introducer Kit for ImpellaRecalls & reported problems across all models

Oscor Inc. · 11 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

9 open recalls name at least one Introducer Kit for Impella model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1,326
Deaths reported
234
Injuries reported
697
Recalls
9

We found 11 models of Introducer Kit for Impella by Oscor Inc. in the FDA's device registry (GUDID). Together they are named in 1,326 adverse event reports in the FDA's MAUDE database, received between Jul 10, 2017 and Sep 30, 2026 — classified as 234 deaths, 697 injuries, 395 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (76 reports).

9 FDA recall records name at least one of these models by its device identifier, 9 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 25 reportsDec 2023: 12 reportsJan 2024: 18 reportsFeb 2024: 3 reportsMar 2024: 35 reportsApr 2024: 84 reportsMay 2024: 46 reportsJun 2024: 38 reportsJul 2024: 16 reportsAug 2024: 77 reportsSep 2024: 69 reportsOct 2024: 361 reportsNov 2024: 29 reportsDec 2024: 10 reportsJan 2025: 23 reportsFeb 2025: 12 reportsMar 2025: 21 reportsApr 2025: 15 reportsMay 2025: 28 reportsJun 2025: 27 reportsJul 2025: 17 reportsAug 2025: 9 reportsSep 2025: 21 reportsOct 2025: 12 reportsNov 2025: 4 reportsDec 2025: 31 reportsJan 2026: 26 reportsFeb 2026: 4 reportsMar 2026: 11 reportsApr 2026: 7 reportsMay 2026: 5 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 5 reportsSep 2026: 2 reportsOct 2026: 0 reports202420252026361
Last 36 months. The most recent month may be incomplete.

Introducer Kit for Impella recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 11, 2026Open, Classified

    A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

    Z-2690-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Jun 10, 2026Open, Classified

    Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.

    Z-2680-2026 · Root cause: Device Design · Recall details · FDA record ↗

  3. Jun 10, 2026Open, Classified

    Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.

    Z-2679-2026 · Root cause: Device Design · Recall details · FDA record ↗

  4. Jun 10, 2026Open, Classified

    Due to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.

    Z-2678-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  5. Aug 5, 2024Open, Classified

    A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

    Z-2955-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  6. Dec 27, 2023Open, Classified

    IFU has been updated to include warnings about the risk of the inlet perforating through the myocardial wall of the left ventricle due to operator handling.

    Z-0980-2024 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  7. Dec 27, 2023Open, Classified

    New warnings are being added to the Instructions for Use for all affected products to reduce the possibility of fibers being drawn into the Impella, which may result in low flow of the device.

    Z-1074-2024 · Root cause: Labeling Change Control · Recall details · FDA record ↗

  8. Nov 7, 2023Open, Classified

    Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis.

    Z-0649-2024 · Root cause: Process control · Recall details · FDA record ↗

  9. Nov 7, 2023Open, Classified

    Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis

    Z-0646-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported Introducer Kit for Impella problems

Device problems

  • Adverse Event Without Identified Device or Use Problem757
  • Fluid/Blood Leak76
  • Pumping Stopped61
  • Deformation Due to Compressive Stress59
  • Patient-Device Incompatibility58
  • Restricted Flow rate58
  • Increase in Pressure39
  • Device Sensing Problem38

Patient problems

  • No Clinical Signs, Symptoms or Conditions386
  • Hemorrhage/Blood Loss/Bleeding268
  • Hemolysis195
  • Ischemia101
  • Thrombosis/Thrombus90
  • Stroke/CVA83
  • Hematoma77
  • Thrombocytopenia39

Introducer Kit for Impella models

11 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Introducer Kit for Impella adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Oscor Inc. products

All Oscor Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page