Arterial Line Kit recallZ-2708-2026 · Argon Medical Devices, Inc · initiated Jun 1, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Argon Medical Devices, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Kits containing Lidocaine HCL Injection USP 1% 10mg/mL 5mL ampules subject to another recall recalled due to not meeting release criteria for sterility.
FDA root cause category: Material/Component Contamination
What the company is doing
On 6/1/2026, correction notices were sent to customers who were asked to do the following: 1) Do not use the affected Lidocaine vials included in affected Kits and Sets in scope of this notice 2) Review your inventory for any product covered by this Medical Device Notification. If any items from the affected lot are identified: - Place a copy of this notice with the affected product - Apply the enclosed stickers to the packaging (kit) to clearly identify the item and remind users to discard affected lidocaine. 3) At the point of use, upon opening the kit, discard the affected Lidocaine vials in accordance with your facility's procedures. 4) Obtain Lidocaine separately through your facility pharmacy and prepare / administer it per institutional protocols prior to the procedure. 5) Complete and return the response form via email [email protected]
Recall details
- Recall number
- Z-2708-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 1, 2026
- Posted by the FDA
- Jul 10, 2026
- Recalling firm
- Argon Medical Devices, Inc
- Quantity
- 440
- Product code
- GDY
- FDA submissions
- K780871
- Distribution
- Distribution US nationwide including Puerto Rico; OUS to Trinidad and Tobago.
Product description
Arterial Line Kit 20ga x 6, REF: 400115A
Code information (models, lots, serial numbers)
REF/UDI-DI Box:Unit/Lot(Expiration): 400115A/20886333212884:00886333212880/11655174(7/19/2027), 11655183(7/19/2027), 11655185(7/19/2027), 11659131(7/19/2027);
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Arterial Line Kit 400115AArgon Medical Devices, Inc. · Gauze/Sponge, Internal, X-Ray Detectable1 recall0 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K780871 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Arterial Line Kit/1 498100Argon Medical Devices, Inc. · Gauze/Sponge, Internal, X-Ray Detectable3 reports
- Arterial Line Kit/1 498106Argon Medical Devices, Inc. · Gauze/Sponge, Internal, X-Ray Detectable3 reports
- Arterial Catheter Mini-Kit 498101Argon Medical Devices, Inc. · Gauze/Sponge, Internal, X-Ray Detectable2 reports
- Arterial Line Kit 498411BArgon Medical Devices, Inc. · Gauze/Sponge, Internal, X-Ray Detectable2 reports
- Arterial Line Kit/1 498104Argon Medical Devices, Inc. · Gauze/Sponge, Internal, X-Ray Detectable1 report
- Arterial Catheter Mini-Kit 498107Argon Medical Devices, Inc. · Gauze/Sponge, Internal, X-Ray Detectable1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 99069).
- Z-2704-2026 · Open, ClassifiedTransvenous Pacing Electrode Tray w/ Bipolar Balloon Pacing Electrode 5F, REF: 008416A, 008566A
- Z-2705-2026 · Open, ClassifiedOne-Piece Introducer Tray, REF: 497455, 497434 cardiovascular
- Z-2706-2026 · Open, ClassifiedGeneral Biospy Tray, REF: GUTS1000
- Z-2707-2026 · Open, ClassifiedT-Lok Bone Marrow Tray, REF: BMT1104STL; BMET1104TL; BMT1104CATL
- Z-2709-2026 · Open, ClassifiedBone Marrow Tray, REF: CB0033TL; CB0355TLA
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page