EndoFLIP recallZ-2950-2026 · Covidien, LLC · initiated May 13, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Covidien, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to products are incorrectly flagged as expired due to the system's 24-hour usage limit, rendering the product unavailable for clinical procedures.

FDA root cause category: Under Investigation by firm

What the company is doing

On 05/13/2026, an "URGENT: MEDICAL DEVICE CORRECTION" letter was emailed/mailed to inform customer that EF-100 modules are affected by an end-of-life (EOL) software issue that prevents the units from recognizing Endoflip and Esoflip catheters after 01-January-2026, causing them to appear as expired. On 07/145/2026, an "Update to Urgent: Medical Device Correction" letter was sent via UPS 2-Day delivery updated potential patient harms Customers are instructed that: "May use the EF-300 system, if they have one. It is recommended to consult their biomedical staff for assistance in adjusting the internal clock. Instructions for this process are enclosed with this notification. As the product has reached end-of-life and end-of-service, Medtronic Technical Services is unable to assist in resetting the internal clock. The user is advised that effects of the data change workaround are seen in exported files. When a USB drive is inserted into the EF-100 system, the SAVE button is enabled to record raw data. For each procedure, a folder is generated and is automatically deleted when the next procedure begins. The folder name includes the Patient ID, the date (YYMMDD format), and the time (HH:MM). All files generated within this folder include the Patient ID in their filenames. The live data file contains raw data with time stamps. These timestamps will reflect the system date, which, under the Date Change Workaround, will be set to 2020. "Post this notice in all care environments where the affected products are used, to maintain awareness to this issue. "Share this notice with all relevant personnel within your organization, and with any organization to which the affected product listed in the Product Information table above has been transferred or distributed. "Complete and return the enclosed Customer Confirmation Form to [insert e-mail], even if you do not have any affected product in your inventory, to acknowledge that you have read and understand this letter.

Recall details

Recall number
Z-2950-2026
Status
Open, Classified
Initiated by the company
May 13, 2026
Posted by the FDA
Aug 14, 2026
Recalling firm
Covidien, LLC
Quantity
327
Product code
FFX
FDA submissions
K183072
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Costa Rica, Denmark, France, Germany, Hong Kong, Ireland, Israel, Italy, Netherlands, New Zealand, Northern Ireland, Norway, Portugal, South Korea, Spain, Sweden, Switzerland, United Kingdom.

Product description

EndoFLIP System that includes: ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307 REFURBISHED SYSTEM RFG-EF100 ENDOFLIP 1 10884521808096 05391530810166 DL ENDOFLIP 1 10884521786684

Code information (models, lots, serial numbers)

ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307 REFURBISHED SYSTEM RFG-EF100 ENDOFLIP 1 10884521808096 05391530810166 DL ENDOFLIP 1 10884521786684