Skip to content

EndoFLIP EF-100 (UDI 10884521809307)Recalls & reported problems

Covidien LP · ENDOFLIP 1 SYSTEM EF-100 · Class 2

All models of EndoFLIP (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521809307), received between Jan 23, 2023 and Oct 15, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Smoking” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Smoking3
  • Adverse Event Without Identified Device or Use Problem1
  • Failure to Power Up1
  • Intermittent Continuity1
  • Thermal Decomposition of Device1
  • Unintended Electrical Shock1

Patient problems

  • Electric Shock1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2025Thermal Decomposition of Device; Failure to Power Up; Smoking; Thermal Decomposition of Device; Failure to Power Up; Smoking

    IT WAS REPORTED THAT DURING THE PROCEDURE, BETWEEN THE FIRST AND SECOND MEASUREMENT, THE SYSTEM'S MODULE SMOKED, A SMELL OCCURRED AND THEN IT TURNED OFF. IT SEEMED A COMPONENT HAD BLOWN. THE SURGEON PULLED OUT THE CATHETER AND CONTINUED THE SURGERY WITHOUT THE DEVICE. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 23277819 ↗
  • MalfunctionApr 25, 2024Intermittent Continuity; Smoking; Intermittent Continuity; Smoking

    ACCORDING TO THE REPORTER, THE UNIT BEEPS WHEN POWERED ON. THE FAN COMES ON AND IT CONTINUED TO BEEP. CUSTOMER TRIED DIFFERENT POWER PACKS AND RESTARTED AND THE UNIT JUST DID NOT WORK. THERE WAS A SMOKE SEEN COMING FROM THE UNIT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19182714 ↗
  • MalfunctionJan 23, 2023Smoking; Adverse Event Without Identified Device or Use Problem; Unintended Electrical Shock; Smoking; Adverse Event Without Identified Device or Use Problem; Unintended Electrical Shock

    ACCORDING TO THE REPORTER, THE UNIT WAS FLIPPED ON AND THE UNIT ITSELF SPARKED AND STARTED SMOKING FROM UNDER THE MACHINE BELOW THE ON/OFF TOGGLE. THE CUSTOMER WAS SHOCKED BY THE DEVICE. THE DEVICE WAS NOT SERVICED OR OPENED PRIOR TO THE INCIDENT. THE SHOCK WAS FROM THE MACHINE, THE POWER BRICK WAS NOT INVOLVED.

    Patient problems: Electric Shock

    MAUDE report 16225545 ↗

Other SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL) devices

All SYSTEM, GASTROINTESTINAL MOTILITY (ELECTRICAL) devices →

Sources and updates