HOFFMANN recallZ-2831-2026 · Stryker GmbH · initiated Jul 1, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
FDA root cause category: Device Design
What the company is doing
Firm began notifying consignees on July 1, 2026. Customers were instructed to immediately check their internal inventory to locate affected product. Customers were informed that until further notice, the subject devices are limited to single use only. Following first use, discard the devices in accordance with the facility's established procedures. Users should conduct visual inspection of product prior to each use. If cracking is found to be present, the customer should quarantine non-confirming parts and report the issue to Stryker's Trauma & Extremities division. Rods with cracks should be returned as part of the complaint process.
Recall details
- Recall number
- Z-2831-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 1, 2026
- Posted by the FDA
- Jul 29, 2026
- Recalling firm
- Stryker GmbH
- Quantity
- 9,405 units
- Product code
- KTT
- FDA submissions
- K033476
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, ID, IL, KS, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, PA, RI, SC, TN, TX, VA, WI, WV and the countries of Australia, Austria, Belgium, Canada, Chile, Denmark, France, Germany, Greece, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Poland, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, UK.
Product description
Stryker Hoffmann II Connecting Rod Micro 3 x 90mm. Catalog Number: 50796090.
Code information (models, lots, serial numbers)
Catalog Number: 50796090. GTIN: 04546540389930. All lot numbers.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HOFFMANN 5079-6-090Stryker GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 recall16 reports
Product lines
Other devices under the same FDA submission
5 devices cleared under K033476 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- HOFFMANN 5079-6-120Stryker GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 recall23 reports
- HOFFMANN 5079-6-050Stryker GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 recall14 reports
- HOFFMANN 5079-6-060Stryker GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 recall12 reports
- HOFFMANN 5079-6-150Stryker GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 recall10 reports
- HOFFMANN 5079-6-030Stryker GmbH · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT1 recall5 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99246).
- Z-2827-2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 30mm. Catalog Number: 50796030.
- Z-2828-2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 40mm. Catalog Number: 50796040.
- Z-2829-2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 50mm. Catalog Number: 50796050.
- Z-2830-2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 60mm. Catalog Number: 50796060.
- Z-2832-2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 120mm. Catalog Number: 50796120.
- Z-2833-2026 · Open, ClassifiedStryker Hoffmann II Connecting Rod Micro 3 x 150mm. Catalog Number: 50796150.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page