Optima recallZ-2726-2026 · GE Medical Systems, LLC · initiated Jun 5, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.
FDA root cause category: Process control
What the company is doing
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 6/5/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use the system by following the instructions below: Perform a check of your device prior to each patient exam to look for the following behaviors: "Column vertical movement requires increased effort. "Column has change in resistance (i.e. movement is not smooth) during vertical movement. "Column makes abnormal noises during vertical movement. If any are observed, DO NOT use the device and call GE HealthCare Service Immediately. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 10967 Digital Response Form or print, fill out manually, scan, and email to [email protected].
Recall details
- Recall number
- Z-2726-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 5, 2026
- Posted by the FDA
- Jul 13, 2026
- Recalling firm
- GE Medical Systems, LLC
- Quantity
- 4 units
- Product code
- IZL
- FDA submissions
- K142383, K103476
- Distribution
- Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
Product description
Optima XR200amx, Mobile Digital-Ready Radiographic System
Code information (models, lots, serial numbers)
UDI-DI: 00840682115278; System IDs: 419383DP3, 773792XR200, 856355OPT, FEB520100
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Optima XR200amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls1 report
Product lines
Other devices under the same FDA submission
1 device cleared under K142383, K103476 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Optima XR220amxGE Medical Systems, L.L.C. · System, x-ray, mobile2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99279).
- Z-2727-2026 · Open, ClassifiedOptima XR220amx, Mobile Digital-Ready Radiographic System
- Z-2728-2026 · Open, ClassifiedOptima XR240amx, Mobile Digital Radiographic System
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page