Skip to content

Astral recallZ-2735-2026 · ResMed Ltd. · initiated Jun 25, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ResMed Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.

FDA root cause category: Component design/selection

What the company is doing

ResMed Urgent Medical Device Correction Astral 100/150 Ventilators Issue Identified: On June 25, 2026, ResMed notified customers via Salesforce of an urgent medical device correction involving a subset of Astral 100 and Astral 150 ventilators. An internal supercapacitor may leak electrolyte over time, potentially damaging the printed circuit board assembly (PCBA) and causing the ventilator to enter a fail-safe state. Interim Guidance: Due to limited PCBA replacement availability, immediate correction of all affected units is not feasible. Physicians and patients are advised to follow Astral 100/150 User and Clinical Guides, including recommendations on monitoring, emergency preparedness, caregiver training, and backup ventilation. Required Actions for Healthcare Providers and Distributors: Acknowledge: Complete and return the acknowledgment form by July 31, 2026. Notify: Distribute the correction notice and accompanying letters to all relevant healthcare professionals, patients, and caregivers immediately. Reinforce Safety Measures: Ensure ventilator-dependent patients are monitored, caregivers are trained in alarm response and emergency procedures, and backup ventilation equipment is functional and readily available. Device Management: Do not remove an affected ventilator from use unless an alternative is immediately available. Replace inoperable units promptly; patients should continue therapy unless directed otherwise by their clinician. Risk Assessment: Evaluate affected patients by clinical risk category (Tier 1, 2, or 3) per Appendix A. Arrange Inspection: Schedule ventilator inspections per the Phase 1 Inspection and Correction Strategy in Appendix A, returning units to an authorized service center as needed. Contact: Reach your local ResMed representative or visit www.resmed.com/contact for questions.

Recall details

Recall number
Z-2735-2026
Status
Open, Classified
Initiated by the company
Jun 25, 2026
Posted by the FDA
Jul 28, 2026
Recalling firm
ResMed Ltd.
Quantity
168,069 devices globally
Product code
CBK
FDA submissions
K152068, K172875
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Israel, Italy, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saint Martin (French part), Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Viet Nam.

Product description

Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.

Code information (models, lots, serial numbers)

Catalog Number: Astral 100 UDI-DI code: 00619498003259 Serial Numbers: Astral Ventilators-Serial number < 22241890149 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070 Catalog Number: Astral 150 UDI-DI code: 00619498270033 Serial Numbers: Astral Ventilator-Serial numbers < 22241440282 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page