Skip to content

Brand Name: COAPTIUM CONNECT recallZ-2903-2026 · Tissium SA · initiated Jun 15, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Tissium SA or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

FDA root cause category: Incorrect or no expiration date

What the company is doing

On June 25, 2026, the firm began notifying consignees via direct visit by medical representatives to deliver the Customer Recall Notification Letter in person. Representatives removed affected units from customer shelves, collected completed response forms, and provided conforming replacement product on the same day.

Recall details

Recall number
Z-2903-2026
Status
Open, Classified
Initiated by the company
Jun 15, 2026
Posted by the FDA
Aug 6, 2026
Recalling firm
Tissium SA
Product code
SFD
Distribution
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.

Product description

Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No

Code information (models, lots, serial numbers)

Lot Code: Model No NCOCO2010; UDI-DI 03760279380094; Lot Number 250700042, 251000033, and 251000034

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 99345).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page