Brand Name: COAPTIUM CONNECT recallZ-2908-2026 · Tissium SA · initiated Jun 15, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Tissium SA or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
FDA root cause category: Incorrect or no expiration date
What the company is doing
On June 25, 2026, the firm began notifying consignees via direct visit by medical representatives to deliver the Customer Recall Notification Letter in person. Representatives removed affected units from customer shelves, collected completed response forms, and provided conforming replacement product on the same day.
Recall details
- Recall number
- Z-2908-2026
- Status
- Open, Classified
- Initiated by the company
- Jun 15, 2026
- Posted by the FDA
- Aug 6, 2026
- Recalling firm
- Tissium SA
- Product code
- SFD
- Distribution
- US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Product description
Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO6010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair device Component: No
Code information (models, lots, serial numbers)
Lot Code: Model No NCOCO6010; UDI-DI ¿03760279380209; Lot Numbers 250700051, 251000044, and 251000045
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- COAPTIUM® Connect NCOCO6010Tissium · In Situ Polymerizing Peripheral Nerve Repair Device1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 99345).
- Z-2901-2026 · Open, ClassifiedBrand Name: COAPTIUM Connect Product Name: COAPTIUM Connect Model/Catalog Number: NCOCO1510 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair …
- Z-2902-2026 · Open, ClassifiedBrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO1575 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair …
- Z-2903-2026 · Open, ClassifiedBrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair …
- Z-2904-2026 · Open, ClassifiedBrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO2075 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair …
- Z-2905-2026 · Open, ClassifiedBrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO3010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair …
- Z-2906-2026 · Open, ClassifiedBrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO4010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair …
- Z-2907-2026 · Open, ClassifiedBrand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Software Version: N/A Product Description: In situ polymerizing peripheral nerve repair …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page