Skip to content

Merge PACS recallZ-2875-2026 · Merge Healthcare, Inc. · initiated Jun 30, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Merge Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient study is open.

FDA root cause category: Software Design Change

What the company is doing

Merge Healthcare issued an Urgent Field Safety Notice (Revision 2) to its consignees on 6/30/2026 via email. The notice explained the issue and when it may occur, risk to patient, and requested an immediate workaround, a configuration change, be performed to disable automatic refresh of the Thumbnail Bar until a software update is available. The notice also directions the consignee to forward the notice to any individuals within their organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. The notice was followed by telephone calls beginning on 7/6/2026 to ensure the notice was received. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge.

Recall details

Recall number
Z-2875-2026
Status
Open, Classified
Initiated by the company
Jun 30, 2026
Posted by the FDA
Aug 4, 2026
Recalling firm
Merge Healthcare, Inc.
Quantity
5 units
Product code
LLZ
FDA submissions
K192455
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, and UK.

Product description

Merge OrthoPACS, medical imaging diagnostic workstation

Code information (models, lots, serial numbers)

UDI/DI (01)00842000101154(10)9.4.0.P02(11)260302, with Affected Versions: 9.4, 9.4 P01, and 9.4 P02.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K192455 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99391).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page