Merge PACS 9.2Recalls & reported problems
Merge Healthcare Solutions Inc. · Class 2
All models of Merge PACS (9)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842000101154), received between Mar 22, 2025 and Mar 22, 2025 — classified as 1 injury.
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jun 30, 2026Open, Classified
Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient study is open.
Z-2874-2026 · Root cause: Software Design Change · Recall details · FDA record ↗
- Jun 30, 2026Open, Classified
Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient study is open.
Z-2875-2026 · Root cause: Software Design Change · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MERGE HEALTHCARE IS SUBMITTING THIS REPORT TO US FDA DUE TO AN ISSUE WITH A THIRD-PARTY APPLICATION USED FOR REPORTING AND DICTATION THAT LED TO A POTENTIAL PATIENT SAFETY RISK. THE THIRD-PARTY APPLICATION INTEROPERATES WITH THE COMPANY'S MERGE PACS AND MERGE WORKFLOW ORCHESTRATOR PRODUCTS. NEITHER THE THIRD-PARTY APPLICATION NOR MERGE WORKFLOW ORCHESTRATOR ARE REGULATED AS MEDICAL DEVICES. A CUSTOMER CONTACTED MERGE HEALTHCARE TECHNICAL SUPPORT ON (B)(6)2025 TO REPORT THAT A PATIENT'S IMAGING STUDY HAD BEEN READ INCORRECTLY BY A RADIOLOGIST. THE CUSTOMER ALLEGED THAT MERGE PACS AND THE DICTATION SYSTEM DISPLAYED IMAGES FOR THE INCORRECT PATIENT. AS A RESULT, A PATIENT WAS SENT OFFSITE FOR FURTHER TREATMENT. NO PATIENT HARM OCCURRED, AS HOSPITAL STAFF IDENTIFIED THAT THE CLINICAL HISTORY IN THE REPORT DID NOT MATCH THE ORDER REQUISITION FOR THE PATIENT. MERGE HEALTHCARE INITIATED AN INVESTIGATION WITH THE THIRD-PARTY VENDOR FOR THE DICTATION SYSTEM. BASED ON AN ANALYSIS OF THE SYSTEM LOG FILES, IT WAS DETERMINED THAT THE USER HAD OPENED TWO PATIENT EXAMS IN QUICK SUCCESSION. THE RESULTS OF THE INVESTIGATION CONCLUDED THAT THE CORRECT IMAGES FOR THE PATIENT WERE DISPLAYED BY MERGE P
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21671425 ↗
Other System, Image Processing, Radiological devices
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Synapse Radiology PACS Synapse Radiology PACS Software LicenseFujifilm Healthcare Americas Corporation · System, image processing, radiological9 recalls38 reports
- CENTRICITY UNIVERSAL VIEWER 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER ZFP CLIENT 6.0Ge Healthcare · System, image processing, radiological8 recalls0 reports
- CENTRICITY UNIVERSAL VIEWER 7.0Ge Healthcare · System, image processing, radiological6 recalls0 reports
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- VMA 2Ortho Kinematics, Inc. · System, Image Processing, Radiological4 recalls0 reports
- CA1000 2.0 SP 14Ge Healthcare · System, image processing, radiological4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.