Merge PACS 9.2Recalls & reported problems

Merge Healthcare Solutions Inc. · Class 2

All models of Merge PACS (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00842000101154), received between Mar 22, 2025 and Mar 22, 2025 — classified as 1 injury.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 30, 2026Open, Classified

    Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient study is open.

    Z-2874-2026 · Root cause: Software Design Change · Recall details · FDA record ↗

  2. Jun 30, 2026Open, Classified

    Merge Healthcare has identified a potential issue that may cause thumbnails from a previously opened patient study to be displayed in the Thumbnail Bar while a different patient study is open.

    Z-2875-2026 · Root cause: Software Design Change · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MERGE HEALTHCARE IS SUBMITTING THIS REPORT TO US FDA DUE TO AN ISSUE WITH A THIRD-PARTY APPLICATION USED FOR REPORTING AND DICTATION THAT LED TO A POTENTIAL PATIENT SAFETY RISK. THE THIRD-PARTY APPLICATION INTEROPERATES WITH THE COMPANY'S MERGE PACS AND MERGE WORKFLOW ORCHESTRATOR PRODUCTS. NEITHER THE THIRD-PARTY APPLICATION NOR MERGE WORKFLOW ORCHESTRATOR ARE REGULATED AS MEDICAL DEVICES. A CUSTOMER CONTACTED MERGE HEALTHCARE TECHNICAL SUPPORT ON (B)(6)2025 TO REPORT THAT A PATIENT'S IMAGING STUDY HAD BEEN READ INCORRECTLY BY A RADIOLOGIST. THE CUSTOMER ALLEGED THAT MERGE PACS AND THE DICTATION SYSTEM DISPLAYED IMAGES FOR THE INCORRECT PATIENT. AS A RESULT, A PATIENT WAS SENT OFFSITE FOR FURTHER TREATMENT. NO PATIENT HARM OCCURRED, AS HOSPITAL STAFF IDENTIFIED THAT THE CLINICAL HISTORY IN THE REPORT DID NOT MATCH THE ORDER REQUISITION FOR THE PATIENT. MERGE HEALTHCARE INITIATED AN INVESTIGATION WITH THE THIRD-PARTY VENDOR FOR THE DICTATION SYSTEM. BASED ON AN ANALYSIS OF THE SYSTEM LOG FILES, IT WAS DETERMINED THAT THE USER HAD OPENED TWO PATIENT EXAMS IN QUICK SUCCESSION. THE RESULTS OF THE INVESTIGATION CONCLUDED THAT THE CORRECT IMAGES FOR THE PATIENT WERE DISPLAYED BY MERGE P

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21671425 ↗

Other System, Image Processing, Radiological devices

All System, Image Processing, Radiological devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page