Midmark IQecg Digital ECG recallZ-3012-2026 · Midmark Corporation · initiated Jul 20, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Midmark Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.
FDA root cause category: Software design
What the company is doing
n URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/17/26 was sent to customers. Safety Instructions Until the corrected IQecg Plugin software has been installed, for electrocardiogram tracings that reflect a fast heart rate (tachycardia), Midmark recommends performing a manual clinical assessment of the patient's heart rate prior to initiating therapeutic intervention. This bedside verification should be used to confirm the accuracy of the displayed ECG heart rate before making clinical management decisions based on the device's displayed value. If this issue is encountered, follow the temporary workaround instructions provided in Appendix A. Midmark will contact your facility with additional information regarding verification of your software installation and implementation of this Field Corrective Action. Product Correction Midmark has developed a corrected version of the IQecg Plugin software that addresses the issue described in this notification and is implementing a Field Corrective Action for affected software versions. No immediate action is required to contact Midmark at this time. Midmark will contact your facility to verify the IQecg Plugin software version(s) installed on your workstations and coordinate the appropriate correction process. Depending on your facility's configuration and support requirements, the correction may be completed using Midmark-provided installation instructions or with assistance coordinated by Midmark Technical Support or an authorized service provider. Contact Information Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact the Technical Service Team at 1.844.856.1230, option 2, or email [email protected].
Recall details
- Recall number
- Z-3012-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 20, 2026
- Posted by the FDA
- Aug 25, 2026
- Recalling firm
- Midmark Corporation
- Quantity
- 3624 units
- Product code
- DPS
- FDA submissions
- K103640
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
Product description
Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.
Code information (models, lots, serial numbers)
Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark IQecg Digital ECG. The Midmark IQecg Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 480000 4820704 UDI-DI: 00841709100581, 00841709100321, 00841709100895, 00841709100901
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Midmark IQecg® Digital ECG 1-100-1330Midmark Corporation · Electrocardiograph1 recall1 report
- Midmark IQecg® Digital ECG 4-000-0062Midmark Corporation · Electrocardiograph1 recall0 reports
- Midmark IQecg® Digital ECG 4-000-0061Midmark Corporation · Electrocardiograph1 recall0 reports
- Midmark IQecg® Digital ECG 1-100-1325Midmark Corporation · Electrocardiograph1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99404).
- Z-3013-2026 · Open, ClassifiedMidmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page