McKesson CONSULT hCG Urine Cassette, Model/ Part:… recallZ-3271-2026 · Alere San Diego, Inc. · initiated Aug 20, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Alere San Diego, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
FDA root cause category: Device Design
What the company is doing
On 8/20/2026, correction notices started being emailed to customers and distributors informing them to take the following actions: 1) In situations where an unexpected test result is obtained, consider all limitations, including peri-menopause that may result in low levels of hCG. A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated, and an alternate confirmatory test that supports a timely clinical decision has been considered. 2) If you have forwarded affected products to other laboratories within your distribution network, notify them of this Field Corrective Action and provide them with a copy of this letter. 3) Retain this letter for your records. 4) Acknowledge receipt of this letter. Your organization s reply provides us with the necessary documentation to monitor the effectiveness of the field action notification. If you have questions regarding this information, U.S. Customers please contact the firm's Technical Services at 877-308-8289
Recall details
- Recall number
- Z-3271-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 20, 2026
- Posted by the FDA
- Sep 25, 2026
- Recalling firm
- Alere San Diego, Inc.
- Quantity
- 50,265,835
- Product code
- JHI
- FDA submissions
- K062361
- Distribution
- Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ireland, Spain, Portugal, New Zealand, Hong Kong, Australia, Papua New Guinea, Fiji, Guam, Sweden.
Product description
McKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine, Model/ Part: FHC-A102/92210; Fisher Scientific Sure-Vue Serum/UrinehCG-STAT, Model/ Part: FHC-A202/23900531, FHC-A202/23900530; McKesson CONSULT Diagnostics hCG Urine Cassette, Model/ Part: FHC-A202/5002; Alere hCG Combo Cassette (20/10 mIU/mL) for Urine or Serum, Model/ Part: FHC-A202/92215; Cardinal Health hCG Combo Rapid Test Serum or Urine, Model/ Part: FHC-A202/B-1077-23; Medline hCG Combo Pregnancy Test Cassette, Model/ Part: FHC-A202/MPH22025HCG;
Code information (models, lots, serial numbers)
All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 5001/ 20612479202744/1155958203, 12/15; 92210/10815845020130/1156015803, 07/16; 23900530/00613647000571/1155810105, 2018-01-15; 23900531/00613647000533/1155810105, 2018-01-15; 5002/20612479202768/1155958305, 12/15; 92215/10815845020123/1156015903, 07/16; B-1077-23/20885380020268/1155904805, 9/2014; MPH22025HCG/10884389191880/1156093402, 04-01-2015
Devices named in this recall
8 devices in the FDA device registry whose identifier appears in the recall's code information.
- McKesson 5001Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls165 reports
- CARDINAL HEALTH B1077-23Cardinal Health 200, LLC · Visual, pregnancy hCG, prescription use1 recall149 reports
- Sure-Vue 4581405021Fisher Scientific Company L.L.C. · Visual, pregnancy hCG, prescription use1 recall83 reports
- McKesson 5002Mckesson Medical-Surgical Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls33 reports
- AlereTM 92210Alere San Diego, Inc. · Visual, Pregnancy Hcg, Prescription Use2 recalls23 reports
- MEDLINE MPH22025HCGMedline Industries, Inc. · Visual, pregnancy hCG, prescription use1 recall53 reports
- Sure-Vue 4581395021Fisher Scientific Company L.L.C. · Visual, pregnancy hCG, prescription use1 recall45 reports
- AlereTM 92215Alere San Diego, Inc. · Visual, Pregnancy Hcg, Prescription Use1 recall6 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99409).
- Z-3272-2026 · Open, ClassifiedFisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman Coulter Inc's ICON 25 hCG, Model/ Part: FHC-202/43025; Henry Schein One Step+…
- Z-3273-2026 · Open, ClassifiedFisher Scientific Sure-Vue Urine hCG Strip, Model/ Part: FHC-101/23900527; McKesson CONSULT hCG Urine Test Dipstick CLIA Waived, Model/ Part:FHC-101/5000; Henry Schein One Step+ hC…
- Z-3274-2026 · Open, ClassifiedFisher Scientific Sure-Vue Urine hCG, Model/ Part: FHC-102/23900526; Henry Schein, Inc. One Step + Urine Cassette Test, Model/ Part: FHC-102/9004074, FHC-102/9004073; Alere hCG Cas…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page