Brand Name: Quantum PureFlow recallZ-3328-2026 · Spectrum Medical Srl · initiated Jul 17, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spectrum Medical Srl or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to blood-to-glycol leak in the heat exchanger.
FDA root cause category: Process design
What the company is doing
On 07/17/2026, an "URGENT MEDICAL DEVICE RECALL" letter was sent via email to customers informing them that, Spectrum Medical S.r.l. has received customer complaints reporting a leak between the blood and glycol compartments of certain adult heat exchangers. Actions to be taken by customers: During the procedure, the user should monitor the heat exchanger for leakage, which would be evident by a change in colour in the glycol compartment. Below are images of the heat exchanger if a blood-to-glycol leak was to occur. When operating as intended, the glycol compartment should remain clear. If the colour of the glycol compartment changes, then a leak is present between the blood and glycol compartments. If leakage is noted between the blood and glycol compartments of the heat exchanger, the heat exchanger should be immediately replaced in the circuit with a new device. Please return the defective heat exchanger to Spectrum Medical. The contaminated heater-cooler device should also be immediately removed from clinical use and returned to Spectrum Medical for decontamination and inspection. Do not use the heater-cooler with another patient. If leakage has occurred, please contact a Spectrum Medical representative to coordinate device return. If leakage into the water compartment occurs when using a third-party heater-cooler, the customer should immediately remove the heater-cooler from clinical use and contact the heater-cooler manufacturer. Spectrum Medical suggests that for customers that use the heat exchanger in combination with a third-party heater-cooler to contact the manufacturer of the heater-cooler to pre-emptively discuss the steps that should be taken if the device is exposed to blood contamination. For questions or further assistance - contact Regulatory Affairs Specialist at [email protected]
Recall details
- Recall number
- Z-3328-2026
- Status
- Open, Classified
- Initiated by the company
- Jul 17, 2026
- Posted by the FDA
- Sep 29, 2026
- Recalling firm
- Spectrum Medical Srl
- Quantity
- 3232
- Product code
- DTR
- FDA submissions
- K220110, K223879
- Distribution
- US Nationwide distribution in the states of AL, CA, FL, KS MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WI.
Product description
Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software Version: N/A Product Description: The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 8 l/min. Quantum PureFlow Standard Heat Exchanger is a single use device which allows heating/cooling the blood during extracorporeal circulation. The device is sold sterile in a blister pack in a multiple box of 4 units. Device surfaces in contact with blood are treated with a phosphorylcholine-based biocompatible coating. The device has been designed to be powered by heaters coolers that use water or a glycol-based heat transfer fluid (HTF). Component: N/A
Code information (models, lots, serial numbers)
Lot Code: 18051160301178 (HX55W-S0)
Other devices under the same FDA submission
2 devices cleared under K220110, K223879 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Quantum PureFlow Standard Heat Exchanger High Flow 3/8 W HX55V-S0WSpectrum Medical Srl · Heat-Exchanger, Cardiopulmonary Bypass1 recall3 reports
- Quantum PureFlow Standard Heat Exchanger High Flow 3/8 HX55W-S0Spectrum Medical Srl · Heat-Exchanger, Cardiopulmonary Bypass4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99462).
- Z-3327-2026 · Open, ClassifiedBrand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 W Model/Catalog Number: HX55V-S0W Software Version: N/A Product Description: The Q…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.