American Contract Systems medical procedure kits and trays… recallZ-2989-2026 · American Contract Systems Inc · initiated May 19, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
FDA root cause category: Nonconforming Material/Component
What the company is doing
American Contract Systems issued an "URGENT: MEDICAL DEVICE FIELD CORRECTION" on 5/19/2026 with a revision issued on 6/4/2026 via email. The notice explained the issue, risks to health, and requested the consignee affix a recall label from the template provided to the exterior of the kit. The label states: "URGENT: RECALL NOTIFICATION, Prior to use, remove & discard the Medicos Ritmed Neurological Sponge per ACS Recall FA-2026-034." Customers/end-users receiving the product through a distributor were directed to contact the distributor or sales representative with any questions. Complete the Recall Response Form. Distributors were directed to directed to ensure all end-users are appropriately notified and maintain records of effectiveness. If you have any questions, contact the Regulatory Affairs Manager, Global Products by email: [email protected] or phone at: 470.280.4277 .
Recall details
- Recall number
- Z-2989-2026
- Status
- Open, Classified
- Initiated by the company
- May 19, 2026
- Posted by the FDA
- Aug 20, 2026
- Recalling firm
- American Contract Systems Inc
- Quantity
- 4789 kits
- Product code
- OJG
- Distribution
- US distribution to states of: MN, MO, IA
Product description
American Contract Systems medical procedure kits and trays labeled as: 1) CRANI PACK - 208400, Model Item Number UDCP57AI; 2) SPINE PACK, Model Item Number HSLM53AC; 3) CRANIAL PACK - 205992, Model Item Number MHCP75AH; 4) IMRI CRANI PACK - 213871, Model Item Number ANCI17AO; 5) CRANI PACK - 205942, Model Item Number ANCP30AP; 6) A-P PACK - 205936, Model Item Number ANAP22AP; 7) POSTERIOR PACK - 205960, Model Item Number ANPP15AR; 8) ANTERIOR CERVICAL PACK - 205935, Model Item Number ANAN51AI; 9) LAMINECTOMY PACK - 205957, Model Item Number ANLM40AG.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
Model/Item Number UDCP57AI, UDI/DI 191072216250, Lot Number 42-8388111; Model/Item Number UDCP57AI, UDI/DI 191072216250, Lot Number 42-8474311; Model/Item Number UDCP57AI, UDI/DI 191072216250, Lot Number 42-8521711; Model/Item Number UDCP57AI, UDI/DI 191072216250, Lot Number 42-8547611; Model/Item Number UDCP57AI, UDI/DI 191072216250, Lot Number 42-8672411; Model/Item Number UDCP57AI, UDI/DI 191072216250, Lot Number 42-8775211; Model/Item Number UDCP57AI, UDI/DI 191072216250, Lot Number 42-8810511; Model/Item Number HSLM53AC, UDI/DI 191072238207, Lot Number 42-8447111; Model/Item Number HSLM53AC, UDI/DI 191072238207, Lot Number 42-8548311; Model/Item Number HSLM53AC, UDI/DI 191072238207, Lot Number 42-8630311; Model/Item Number HSLM53AC, UDI/DI 191072238207, Lot Number 42-8785011; Model/Item Number HSLM53AC, UDI/DI 191072238207, Lot Number 42-8742911; Model/Item Number MHCP75AH, UDI/DI 191072211934, Lot Number 42-8455911; Model/Item Number MHCP75AH, UDI/DI 191072211934, Lot Number 42-8498111; Model/Item Number MHCP75AH, UDI/DI 191072211934, Lot Number 42-8538111; Model/Item Number MHCP75AH, UDI/DI 191072211934, Lot Number 42-8601211; Model/Item Number MHCP75AH, UDI/DI 191072211934, Lot Number 42-8764511; Model/Item Number MHCP75AH, UDI/DI 191072211934, Lot Number 42-8823811; Model/Item Number ANCI17AO, UDI/DI 191072232052, Lot Number 42-8384311; Model/Item Number ANCI17AO, UDI/DI 191072232052, Lot Number 42-8525011; Model/Item Number ANCI17AO, UDI/DI 191072232052, Lot Number 42-8611711; Model/Item Number ANCI17AO, UDI/DI 191072232052, Lot Number 42-8705911; Model/Item Number ANCI17AO, UDI/DI 191072232052, Lot Number 42-8798411; Model/Item Number ANCI17AO, UDI/DI 191072232052, Lot Number 42-8831911; Model/Item Number ANCI17AO, UDI/DI 191072232052, Lot Number 42-8854511; Model/Item Number ANCP30AP, UDI/DI 191072232069, Lot Number 42-8497911; Model/Item Number ANCP30AP, UDI/DI 191072232069, Lot Number 42-8537311; Model/Item Number ANCP30AP, UDI/DI 191072232069, Lot Number 42-8673211; Model/Item Number ANCP30AP, UDI/DI 191072232069, Lot Number 42-8756411; Model/Item Number ANCP30AP, UDI/DI 191072232069, Lot Number 42-8818611; Model/Item Number ANCP30AP, UDI/DI 191072232069, Lot Number 42-8854011; Model/Item Number ANCP30AP, UDI/DI 191072232069, Lot Number 42-8869111; Model/Item Number ANAP22AP, UDI/DI 191072214287, Lot Number 42-8497811; Model/Item Number ANAP22AP, UDI/DI 191072214287, Lot Number 42-8588811; Model/Item Number ANAP22AP, UDI/DI 191072214287, Lot Number 42-8666611; Model/Item Number ANAP22AP, UDI/DI 191072214287, Lot Number 42-8730411; Model/Item Number ANAP22AP, UDI/DI 191072214287, Lot Number 42-8785411; Model/Item Number ANAP22AP, UDI/DI 191072214287, Lot Number 42-8833011; Model/Item Number ANAP22AP, UDI/DI 191072214287, Lot Number 42-8854611; Model/Item Number ANAP22AP, UDI/DI 191072214287, Lot Number 42-8869611; Model/Item Number ANPP15AR, UDI/DI 191072214423, Lot Number 42-8345311; Model/Item Number ANPP15AR, UDI/DI 191072214423, Lot Number 42-8379411; Model/Item Number ANPP15AR, UDI/DI 191072214423, Lot Number 42-8547011; Model/Item Number ANPP15AR, UDI/DI 191072214423, Lot Number 42-8654511; Model/Item Number ANPP15AR, UDI/DI 191072214423, Lot Number 42-8746311; Model/Item Number ANPP15AR, UDI/DI 191072214423, Lot Number 42-8801611; Model/Item Number ANAN51AI, UDI/DI 191072232045, Lot Number 42-8401611; Model/Item Number ANAN51AI, UDI/DI 191072232045, Lot Number 42-8611411; Model/Item Number ANAN51AI, UDI/DI 191072232045, Lot Number 42-8611411A; Model/Item Number ANAN51AI, UDI/DI 191072232045, Lot Number 42-8648311; Model/Item Number ANAN51AI, UDI/DI 191072232045, Lot Number 42-8755811; Model/Item Number ANAN51AI, UDI/DI 191072232045, Lot Number 42-8807911; Model/Item Number ANAN51AI, UDI/DI 191072232045, Lot Number 42-8852311; Model/Item Number ANAN51AI, UDI/DI 191072232045, Lot Number 42-8869511; Model/Item Number ANLM40AG, UDI/DI 191072232076, Lot Number 42-8384511; Model/Item Number ANLM40AG, UDI/DI 191072232076, Lot Number 42-8545711; Model/Item Number ANLM40AG, UDI/DI 191072232076, Lot Number 42-8634611; Model/Item Number ANLM40AG, UDI/DI 191072232076, Lot Number 42-8707111; Model/Item Number ANLM40AG, UDI/DI 191072232076, Lot Number 42-8789211; Model/Item Number ANLM40AG, UDI/DI 191072232076, Lot Number 42-8831811; Model/Item Number ANLM40AG, UDI/DI 191072232076, Lot Number 42-8853011; Model/Item Number ANLM40AG, UDI/DI 191072232076, Lot Number 42-8868811.
Devices named in this recall
8 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACS ANPP15ARAmerican Contract Systems, Inc. · Neurological tray2 recalls0 reports
- ACS ANLM40AGAmerican Contract Systems, Inc. · Neurological tray2 recalls0 reports
- ACS MHCP75AHAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS ANAP22APAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS UDCP57AIAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS ANAN51AIAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS ANCI17AOAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
- ACS ANCP30APAmerican Contract Systems, Inc. · Neurological tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99482).
- Z-2990-2026 · Open, ClassifiedAmerican Contract Systems medical procedure kits and trays labeled as: 1) ANTERIOR CERVICAL PACK - 205988, Model Item Number MHAC65AH; 2) SPINE PACK - 206013, Model Item Number …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.