American Contract Systems medical procedure kits and trays… recallZ-2990-2026 · American Contract Systems Inc · initiated May 19, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from American Contract Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
FDA root cause category: Nonconforming Material/Component
What the company is doing
American Contract Systems issued an "URGENT: MEDICAL DEVICE FIELD CORRECTION" on 5/19/2026 with a revision issued on 6/4/2026 via email. The notice explained the issue, risks to health, and requested the consignee affix a recall label from the template provided to the exterior of the kit. The label states: "URGENT: RECALL NOTIFICATION, Prior to use, remove & discard the Medicos Ritmed Neurological Sponge per ACS Recall FA-2026-034." Customers/end-users receiving the product through a distributor were directed to contact the distributor or sales representative with any questions. Complete the Recall Response Form. Distributors were directed to directed to ensure all end-users are appropriately notified and maintain records of effectiveness. If you have any questions, contact the Regulatory Affairs Manager, Global Products by email: [email protected] or phone at: 470.280.4277 .
Recall details
- Recall number
- Z-2990-2026
- Status
- Open, Classified
- Initiated by the company
- May 19, 2026
- Posted by the FDA
- Aug 20, 2026
- Recalling firm
- American Contract Systems Inc
- Quantity
- 160 kits
- Product code
- OJH
- Distribution
- US distribution to states of: MN, MO, IA
Product description
American Contract Systems medical procedure kits and trays labeled as: 1) ANTERIOR CERVICAL PACK - 205988, Model Item Number MHAC65AH; 2) SPINE PACK - 206013, Model Item Number MHSP99AF.
Code information (models, lots, serial numbers)
Model/Item Number MHAC65AH, UDI/DI 191072211910, Lot Number 42-8606211; Model/Item Number MHAC65AH, UDI/DI 191072211910, Lot Number 42-8687611; Model/Item Number MHAC65AH, UDI/DI 191072211910, Lot Number 42-8847511; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8423411; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8470411; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8527611; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8601311; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8731511; Model/Item Number MHSP99AF, UDI/DI 191072214782, Lot Number 42-8795311.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ACS MHAC65AHAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
- ACS MHSP99AFAmerican Contract Systems, Inc. · Orthopedic tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99482).
- Z-2989-2026 · Open, ClassifiedAmerican Contract Systems medical procedure kits and trays labeled as: 1) CRANI PACK - 208400, Model Item Number UDCP57AI; 2) SPINE PACK, Model Item Number HSLM53AC; 3) CRANIAL…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.