5008X HD/HDF with CLiC Blood recallZ-3086-2026 · Fresenius Medical Care Renal Therapies Group, LLC · initiated Aug 10, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Medical Care Renal Therapies Group, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.

FDA root cause category: Component change control

What the company is doing

On 8/10/2026, the firm began notifying consignees via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to 1. Identify any affected product on hand. 2. IMMEDIATELY segregate and discontinue use of affected lot. 3. Contact the call center at 1-888-943-4901 and reference Event 8514, between 8:00 AM 4:30 PM Eastern daily, for return of affected lots and replacement of your stock. 4, Complete and return the included reply form.

Recall details

Recall number
Z-3086-2026
Status
Open, Classified
Initiated by the company
Aug 10, 2026
Posted by the FDA
Sep 2, 2026
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Quantity
18,600
Product code
KOC
FDA submissions
K243505
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.

Product description

5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C

Code information (models, lots, serial numbers)

Lots 25JR01801 25KR01225 25KR01251 25NR01016 25NR01265 UDI-DI 00840861102532

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

4 devices cleared under K243505 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99556).