Skip to content

Novum IQ recallZ-3066-2026 · Baxter Healthcare Corporation · initiated Aug 11, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.

FDA root cause category: Software design

What the company is doing

On August 11, 2026, firm began sending customers Urgent Medical Device Correction delivered via email and USPS first class mail. Customers were advised to use the same compatible syringe model (brand/size) for all syringes required within a Multi-Syringe VOT workflow. Customers were provided with additional instructions for situations where the syringe does not support the intended flow rate and the user receives a prompt to confirm continuation of the infusion at the maximum or minimum supported flow rate for that syringe after a syringe change.

Recall details

Recall number
Z-3066-2026
Status
Open, Classified
Initiated by the company
Aug 11, 2026
Posted by the FDA
Aug 31, 2026
Recalling firm
Baxter Healthcare Corporation
Quantity
4 units
Product code
FRN
FDA submissions
K242390
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.

Product description

Baxter Novum IQ Syringe Pump, Loaner, Model Number 40800BAXUSL

Code information (models, lots, serial numbers)

UDI-DI 00197886002300, all serial numbers

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 99557).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page