Skip to content

Oncology sets with drip chambers recallZ-3158-2026 · ICU Medical, Inc. · initiated May 12, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.

FDA root cause category: Process control

What the company is doing

On 5/12/2026, the firm posted a Information Regarding Reported Black Specks in Certain IV Tubing Set Drip Chambers from ICU Medical document on their website. On 5/18/2026, updated the notices were sent. On 7/28/2026, correction notices were sent to customers asking them to do the following: 1) Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used or further distributed to another location, please ensure that this communication is delivered there. 2) Complete and return the attached Customer Response Form to [email protected] 3) Continue using products as intended and per facility protocol. 4) Report any product quality concerns to ICU Medical Complaint Management. Firm contacts: [email protected] 1-866-261-8806. [email protected] 1-800-258-5361

Recall details

Recall number
Z-3158-2026
Status
Open, Classified
Initiated by the company
May 12, 2026
Posted by the FDA
Sep 11, 2026
Recalling firm
ICU Medical, Inc.
Quantity
107,800
Product code
FPA
FDA submissions
K081361
Distribution
Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR.

Product description

Oncology sets with drip chambers

Code information (models, lots, serial numbers)

All products manufactured on or before 7/4/2026. UDI-DI: 10840619054486, 10887709091914

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

18 devices cleared under K081361 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99058).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page