Skip to content

Triathlon recallZ-3198-2026 · Howmedica Osteonics Corp. · initiated Jul 30, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

FDA root cause category: Process change control

What the company is doing

Firm began notifying customers on July 30, 2026, via FedEx 2-day air. Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.

Recall details

Recall number
Z-3198-2026
Status
Open, Classified
Initiated by the company
Jul 30, 2026
Posted by the FDA
Sep 16, 2026
Recalling firm
Howmedica Osteonics Corp.
Quantity
12,156 US
Product code
KRO
FDA submissions
K223528
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.

Product description

TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Code information (models, lots, serial numbers)

5612-P-411 UDI-DI 07613327601893 Lot 360958 5612-P-413 UDI-DI 07613327602012 Lot 354573

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 99558).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page