Ref:2b5st recallZ-3183-2026 · Stryker Sustainability Solutions · initiated Aug 19, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Sustainability Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reusable access port system may crack or seperate during use.
FDA root cause category: Device Design
What the company is doing
On August 19, 2026, Stryker Sustainability Solutions (SSS)issued a Urgent: Medical Device recall notice to affected consignees via sales representative and US Mail. SSS ask consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Immediately check your internal inventory to locate the affected product. Segregate the affected products, quarantine and discontinue use. 3. Complete the Business Reply Form or use the QR Code on Attachment A. Response is required. a.If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility.
Recall details
- Recall number
- Z-3183-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 19, 2026
- Posted by the FDA
- Sep 16, 2026
- Recalling firm
- Stryker Sustainability Solutions
- Quantity
- 2838 units
- Product code
- NLM
- FDA submissions
- K201511
- Distribution
- US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina , South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the countries of Israel and Canada.
Product description
REF:2B5ST Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/OPTIVIEW TECHNOLOGY & SLEEVE 5MM X 75MM
Code information (models, lots, serial numbers)
UDI: (01)00885825017231/ Lots: 16188647 16210212 16243934 16267300 16285639 16285656 16306188 16306231 16364978 16470903 16521533 16547722 16580202 16599810 16637989 16656930 16665550 17005619 17069758 17113276 17117460 17146892 17172722 17389157
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99634).
- Z-3182-2026 · Open, ClassifiedREF:2B5LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/OPTIVIEW TECHNOLOGY & SLEEVE 5MM X 100MM
- Z-3184-2026 · Open, ClassifiedREF:B11LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 11X100MM
- Z-3185-2026 · Open, ClassifiedREF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM
- Z-3186-2026 · Open, ClassifiedREF:B5LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 5X100MM
- Z-3187-2026 · Open, ClassifiedREF:B5ST Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 5X75MM
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page