Ref:B12lt recallZ-3185-2026 · Stryker Sustainability Solutions · initiated Aug 19, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Sustainability Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reusable access port system may crack or seperate during use.
FDA root cause category: Device Design
What the company is doing
On August 19, 2026, Stryker Sustainability Solutions (SSS)issued a Urgent: Medical Device recall notice to affected consignees via sales representative and US Mail. SSS ask consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Immediately check your internal inventory to locate the affected product. Segregate the affected products, quarantine and discontinue use. 3. Complete the Business Reply Form or use the QR Code on Attachment A. Response is required. a.If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility.
Recall details
- Recall number
- Z-3185-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 19, 2026
- Posted by the FDA
- Sep 16, 2026
- Recalling firm
- Stryker Sustainability Solutions
- Quantity
- 2682 units
- Product code
- NLM
- FDA submissions
- K201511
- Distribution
- US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina , South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the countries of Israel and Canada.
Product description
REF:B12LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 12X100MM
Code information (models, lots, serial numbers)
UDI: (01)00885825020897/ Lots: 16029001 16032878 16038822 16041414 16044000 16048542 16052751 16054119 16062169 16062738 16066292 16074632 16087947 16087952 16090650 16098040 16107246 16111434 16115899 16115909 16120868 16130680 16130701 16135456 16136279 16138395 16140432 16140568 16142544 16142619 16159304 16160563 16172080 16176548 16186738 16186874 16187194 16187248 16187457 16197336 16201786 16205783 16205810 16205823 16206753 16220653 16220918 16221448 16221453
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 99634).
- Z-3182-2026 · Open, ClassifiedREF:2B5LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/OPTIVIEW TECHNOLOGY & SLEEVE 5MM X 100MM
- Z-3183-2026 · Open, ClassifiedREF:2B5ST Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/OPTIVIEW TECHNOLOGY & SLEEVE 5MM X 75MM
- Z-3184-2026 · Open, ClassifiedREF:B11LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 11X100MM
- Z-3186-2026 · Open, ClassifiedREF:B5LT Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 5X100MM
- Z-3187-2026 · Open, ClassifiedREF:B5ST Ethicon Endo-Surgery, Inc. XCEL BLADELESS TROCAR W/STABILITY SLEEVE 5X75MM
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page