Skip to content

StoneSmart Connect recallZ-3188-2026 · Boston Scientific Corporation · initiated Jul 22, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On July 22, 2026, Boston Scientific Corporation issued a medical device correction to affected consignees via FedEx letter, Boston Scientific Corporation requested consignees to take the following actions: 1. Continue to use the affected devices according to the Instructions for Use. 2. Forward this notice to relevant personnel and any facilities to which the impacted device was transferred. 3. Complete and return the enclosed Acknowledgement Form. 4. Begin using the replacement power adaptor upon receipt and dispose of the affected power adaptor in accordance with local requirements.

Recall details

Recall number
Z-3188-2026
Status
Open, Classified
Initiated by the company
Jul 22, 2026
Posted by the FDA
Sep 16, 2026
Recalling firm
Boston Scientific Corporation
Quantity
982 units
Product code
FGB
FDA submissions
K233645
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and the countries of Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon, Norway, Poland, Portugal, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Turkiye, Canada, Hong Kong, Singapore, Taiwan, Argentina, Chile, Japan.

Product description

Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable REF: M0067933000, M006793300H0, M006793300R0, M006793300D0, M006793300Z0, M0067930000, M006793000R0

Code information (models, lots, serial numbers)

All Serial Numbers: REF: M0067933000 (UDI 00191506002510, UDI (00191506021771); REF: M006793300H0 (UDI 00191506002558); REF: M006793300R0 (UDI 00191506002534), UDI (00191506021788); REF: M006793300D0 (UDI 00191506002527); REF: M006793300Z0 (UDI 00191506002541); UPN M0067933100 (UDI 00191506022549)

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page