Skip to content

StoneSmart™ Connect M006793300D0Recalls & reported problems

Boston Scientific Corporation · Console · Class 2

All models of StoneSmart Connect (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 22, 2026Class IIOpen, Classified

    A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.

    Z-3188-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Other Ureteroscope and accessories, flexible/rigid devices

All Ureteroscope and accessories, flexible/rigid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page