Skip to content

VERIFY recallZ-0075-2027 · Steris Corporation · initiated Aug 31, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

FDA root cause category: Under Investigation by firm

What the company is doing

Firm began notifying customers on August 31, 2026 via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to destroy any affected product on hand. The firm will replace any product still within its shelf life.

Recall details

Recall number
Z-0075-2027
Status
Open, Classified
Initiated by the company
Aug 31, 2026
Posted by the FDA
Oct 2, 2026
Recalling firm
Steris Corporation
Quantity
233
Product code
FRC
FDA submissions
K854902
Distribution
Worldwide - US Nationwide distribution.

Product description

VERIFY Biological Steam Test Pack Model/Catalog Number: S3069

Code information (models, lots, serial numbers)

UDI:10724995023864 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K854902 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 99704).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page