Multichem AE recallZ-0205-2027 · Techno-Path Manufacturing Ltd. · initiated Aug 14, 2026
Class II Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Techno-Path Manufacturing Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
FDA root cause category: Under Investigation by firm
What the company is doing
On August 14, 2026, Techno Path Manufacturing Ltd has issued a medical device correction to consignees via email. Techno Path Manufacturing Ltd informed consignees to please verify their inventory against the affected product codes and lot numbers. Techno Path Manufacturing Ltd. has informed the consignees to take the following actions: 1. Immediately place all affected inventory on hold and prevent further distribution. 2. Quarantine all affected product currently under your control. 3. Confirm that there has been no distribution of the impacted lots to customers. 4. Confirm receipt of this notification and provide inventory and distribution details to Technopath Manufacturing Ltd. within 3 business days. 5. Arrange return or destruction of affected inventory in accordance with instructions provided by Technopath Manufacturing Ltd.
Recall details
- Recall number
- Z-0205-2027
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Aug 14, 2026
- Posted by the FDA
- Oct 22, 2026
- Recalling firm
- Techno-Path Manufacturing Ltd.
- Quantity
- 832 units
- Product code
- OHQ
- Distribution
- US Nationwide distribution in the state of Massachusetts.
Product description
Techno-Path Multichem AE Unassayed Level 1 REF:AE621X
Code information (models, lots, serial numbers)
UDI: (01)05391523442800/Lot # 65705261
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Multichem AE AE621XTechno-Path Manufacturing Limited · Multi-Analyte Controls Unassayed1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99724).
- Z-0206-2027 · Open, ClassifiedTechno-Path Multichem AE Unassayed Level 2 REF: AE622X
- Z-0207-2027 · Open, ClassifiedTechno-Path Multichem AE Unassayed Level 3, REF:AE623X
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Nov 27, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Nov 27, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Nov 27, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page