Skip to content

Multichem AE recallZ-0206-2027 · Techno-Path Manufacturing Ltd. · initiated Aug 14, 2026

Class II Open, Classified All recalls posted in October 2026

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Techno-Path Manufacturing Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.

FDA root cause category: Under Investigation by firm

What the company is doing

On August 14, 2026, Techno Path Manufacturing Ltd has issued a medical device correction to consignees via email. Techno Path Manufacturing Ltd informed consignees to please verify their inventory against the affected product codes and lot numbers. Techno Path Manufacturing Ltd. has informed the consignees to take the following actions: 1. Immediately place all affected inventory on hold and prevent further distribution. 2. Quarantine all affected product currently under your control. 3. Confirm that there has been no distribution of the impacted lots to customers. 4. Confirm receipt of this notification and provide inventory and distribution details to Technopath Manufacturing Ltd. within 3 business days. 5. Arrange return or destruction of affected inventory in accordance with instructions provided by Technopath Manufacturing Ltd.

Recall details

Recall number
Z-0206-2027
Recall class
Class II
Status
Open, Classified
Initiated by the company
Aug 14, 2026
Posted by the FDA
Oct 22, 2026
Recalling firm
Techno-Path Manufacturing Ltd.
Quantity
123 units
Product code
OHQ
Distribution
US Nationwide distribution in the state of Massachusetts.

Product description

Techno-Path Multichem AE Unassayed Level 2 REF: AE622X

Code information (models, lots, serial numbers)

UDI:(01)05391523442817/Lot # 65705262

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 99724).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page