PERMA-HAND recallZ-0153-2027 · Ethicon, LLC · initiated Sep 8, 2026
Open, Classified All recalls posted in October 2026
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ethicon, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.
FDA root cause category: Process control
What the company is doing
On 9/8/2026, recall notices were mailed to customers who were asked to do the following: 1) Quarantine affected products. Maintain a copy of this notice with the quarantined product and keep a copy for your records. 2) Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return and forward them this recall notice. 3) Complete and return the Business Reply Form via email to [email protected] 4) Return affected product using the shipping label provided. In addition distributors were told the following: Based on your reported sales data, Ethicon will send notices directly to customers about the product subject to this recall. This notice will include instructions on returning the product for credit reimbursement. Distribute this notice within your organization and to any non-customer organization that received affected product from your facility. If you have additional questions regarding this communication or want to report any product complaints, please contact the firm's Resource Department at 1-877-384-4266.
Recall details
- Recall number
- Z-0153-2027
- Recall class
- Not yet classified by the FDA
- Status
- Open, Classified
- Initiated by the company
- Sep 8, 2026
- Posted by the FDA
- Oct 14, 2026
- Recalling firm
- Ethicon, LLC
- Quantity
- 1764
- Product code
- GAQ
- FDA submissions
- K946173
- Distribution
- Worldwide - US Nationwide distribution in the states of SC, PA, TX, CA, ID, LA, MO, AL, NC, UT, IL, MA, OH, AZ, TN, OK, NE, ME, IN, WI, FL, CO, WY, NH, NY, NC, MN and the country of CA.
Product description
PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H
Code information (models, lots, serial numbers)
UDI-DI:Box/Unit: 30705031047369/10705031047365. Lot: RLBDPJ. Expiration: SEPTEMBER 30, 2026
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- PERMA-HAND N271HEthicon, LLC · SUTURE, NONABSORBABLE, SILK1 recall2 reports
Product lines
Other devices under the same FDA submission
456 devices cleared under K946173 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ETHILON 697GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE223 reports
- ETHIBOND EXCEL X557HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE187 reports
- PERMA-HAND SA86GEthicon, LLC · SUTURE, NONABSORBABLE, SILK120 reports
- ETHILON 698GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE119 reports
- ETHILON 1696GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE118 reports
- VICRYL V960GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall55 reports
- ETHICON M660GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE102 reports
- PERMA-HAND K833HEthicon, LLC · SUTURE, NONABSORBABLE, SILK1 recall46 reports
- ETHILON 699GEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall34 reports
- ETHILON 1669HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE1 recall30 reports
- VICRYL V449GEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID74 reports
- ETHICON M650GEthicon, LLC · SUTURE, NONABSORBABLE, STEEL, MONOFILAMENT AND MULTIFILAMENT, STERILE68 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page