ETHIBOND EXCEL X557HRecalls & reported problems
Ethicon, LLC · Green Braided Polyester, Nonabsorbable Surgical Suture · Class 2 · Implant
All models of ETHIBOND EXCEL (133)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ethicon Ethibond Excel suture is a green braided polyester thread for closing tissue in surgery. Its 187 reports include no deaths and no injuries, and most describe the suture separating or breaking during procedures, sometimes with the needle popping off. Several reports say surgeons switched to another suture to finish the operation. Reports in the last 12 months (4) are down from the prior 12 months (11), and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 187
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 187 adverse event reports that list this exact device (DI 10705031057562), received between Apr 4, 2018 and Nov 14, 2025 — classified as 187 malfunctions.
The most frequently reported specific device problem is “Material Separation” (175 reports).
10 other recalls cover devices cleared under the same FDA submission (K946173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Separation175
- Break11
- Material Frayed1
Patient problems
- No Clinical Signs, Symptoms or Conditions183
- No Consequences Or Impact To Patient4
Recent adverse event reports
The 25 most recent of 187 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 14, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED OVER THE COURSE OF 2 CASES THE RB1 NEEDLE WAS POPPING OFF THE SUTURE AND IT'S NOT A POP OFF SUTURE. THEY WOULDN'T USE ANYMORE SUTURE FROM THAT BOX. GOT ANOTHER BOX FROM ANOTHER HOSPITAL WITH A DIFFERENT LOT NUMBER TO USE FOR THE NEXT FEW WEEKS OF CASES. NO ADVERSE IMPACT TO THE PATIENT/USER. DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23559970 ↗MalfunctionNov 14, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED OVER THE COURSE OF 2 CASES THE RB1 NEEDLE WAS POPPING OFF THE SUTURE AND IT'S NOT A POP OFF SUTURE. THEY WOULDN'T USE ANYMORE SUTURE FROM THAT BOX. GOT ANOTHER BOX FROM ANOTHER HOSPITAL WITH A DIFFERENT LOT NUMBER TO USE FOR THE NEXT FEW WEEKS OF CASES. NO ADVERSE IMPACT TO THE PATIENT/USER. DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23559955 ↗MalfunctionNov 14, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED OVER THE COURSE OF 2 CASES THE RB1 NEEDLE WAS POPPING OFF THE SUTURE AND IT'S NOT A POP OFF SUTURE. THEY WOULDN'T USE ANYMORE SUTURE FROM THAT BOX. GOT ANOTHER BOX FROM ANOTHER HOSPITAL WITH A DIFFERENT LOT NUMBER TO USE FOR THE NEXT FEW WEEKS OF CASES. NO ADVERSE IMPACT TO THE PATIENT/USER. DEVICE IS NOT AVAILABLE FOR RETURN. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23559946 ↗MalfunctionNov 14, 2025Material Frayed; Material Frayed
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. IT WAS REPORTED THAT THE SUTURE FRAYED. THERE IS NO ADVERSE IMPACT ON PATIENT/USER REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23556949 ↗MalfunctionSep 18, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 5 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23088564 ↗MalfunctionSep 18, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 5 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23088395 ↗MalfunctionSep 18, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 5 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23088388 ↗MalfunctionSep 18, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 5 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23088321 ↗MalfunctionSep 18, 2025Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2025 AND SUTURE WAS USED. DURING THE PROCEDURE, IT WAS REPORTED THAT THE SUTURE WAS BROKEN WHEN THE SURGEON ATTEMPTED TO SEW THE TISSUE. TOTAL 5 PRODUCTS OCCURRED SUTURE BREAKAGE ISSUE. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23088302 ↗MalfunctionAug 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON SUTURES PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. A TOTAL 3 PRODUCTS EXPERIENCED THE NEEDLE PULLING OFF ISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22776179 ↗MalfunctionAug 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON SUTURES PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. A TOTAL 3 PRODUCTS EXPERIENCED THE NEEDLE PULLING OFF ISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22775349 ↗MalfunctionAug 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON SUTURES PROCEDURE ON (B)(6) 2025 AND A SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. A TOTAL 3 PRODUCTS EXPERIENCED THE NEEDLE PULLING OFF ISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22775255 ↗MalfunctionAug 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON SUTURES PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. A TOTAL 3 PRODUCTS EXPERIENCED THE NEEDLE PULLING OFF ISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22774929 ↗MalfunctionAug 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON SUTURES PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. A TOTAL 3 PRODUCTS EXPERIENCED THE NEEDLE PULLING OFF ISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22774762 ↗MalfunctionAug 11, 2025Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A TENDON SUTURES PROCEDURE ON AN UNKNOWN DATE AND A SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. A TOTAL 3 PRODUCTS EXPERIENCED THE NEEDLE PULLING OFF ISSUE. CHANGED TO ANOTHER ONE TO COMPLETE THE SURGERY. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22774604 ↗MalfunctionJan 30, 2024Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT MICROSCOPIC RESECTION OF BRAIN TUMORS PROCEDURE ON (B)(6) 2024 AND SUTURE WAS USED. THE SUTURE GOT SPLIT DURING THE SURGERY. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THERE IS NO REPORT ON PATIENT'S INJURY. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18608593 ↗MalfunctionOct 26, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN EXCISION OF BRAIN TUMORS SURGERY ON (B)(6) 2023 AND SUTURE WAS USED. DURING THE PROCEDURE, PROBLEM OF PULLING OFF SUTURE NEEDLE HAPPENED IN THE EXCISION OF BRAIN TUMORS SURGERY. CHANGED ANOTHER NINE TO CONTINUE THE SURGERY, THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18012980 ↗MalfunctionOct 24, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A MICROSCOPIC RESECTION OF BRAIN TUMORS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER SEVEN TO CONTINUE THE SURGERY, AND THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998999 ↗MalfunctionOct 24, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A MICROSCOPIC RESECTION OF BRAIN TUMORS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER SEVEN TO CONTINUE THE SURGERY, AND THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998998 ↗MalfunctionOct 24, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A MICROSCOPIC RESECTION OF BRAIN TUMORS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER SEVEN TO CONTINUE THE SURGERY, AND THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998997 ↗MalfunctionOct 24, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A MICROSCOPIC RESECTION OF BRAIN TUMORS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER SEVEN TO CONTINUE THE SURGERY, AND THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998992 ↗MalfunctionOct 24, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A MICROSCOPIC RESECTION OF BRAIN TUMORS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER SEVEN TO CONTINUE THE SURGERY, AND THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998987 ↗MalfunctionOct 24, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A MICROSCOPIC RESECTION OF BRAIN TUMORS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER SEVEN TO CONTINUE THE SURGERY, AND THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998984 ↗MalfunctionOct 24, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A MICROSCOPIC RESECTION OF BRAIN TUMORS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER SEVEN TO CONTINUE THE SURGERY, AND THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998974 ↗MalfunctionOct 24, 2023Material Separation; Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT A MICROSCOPIC RESECTION OF BRAIN TUMORS PROCEDURE ON AN UNKNOWN DATE AND SUTURE WAS USED. THE PROBLEM OF THE SUTURE PULLING OFF THE NEEDLE HAPPENED IN SURGERY. CHANGED TO ANOTHER SEVEN TO CONTINUE THE SURGERY, AND THE SAME PROBLEM HAPPENED. CHANGED ANOTHER ONE TO COMPLETE THE SURGERY. THE NEEDLE DID NOT FELL INTO THE PATIENT. NO ADVERSE PATIENT CONSEQUENCES WERE REPORTED. ADDITIONAL INFORMATION WAS REQUESTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17998971 ↗
FDA clearance and classification
Cleared through 510(k) K946173 on Jan 9, 1995 (decision: Substantially Equivalent). Submitted by ETHICON, Inc. as “Stainless Steel, Gut, Poly (Glycolide/L-Lactide) Polyamide, Polypropylene, Poly (Ethylene Terephthalate) And Silk”. The FDA review took 21 days. FDA record ↗
The FDA regulates product code GAT (suture, nonabsorbable, synthetic, polyethylene) under 21 CFR 878.5000 as a Class II device. 21 CFR 878.5000 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K946173. One submission often covers many models, so check the product and lot details in each record.
- Z-0284-2027 · Sep 23, 2026 · Open, ClassifiedETHILON Nylon Suture, Product Code 697G
- Z-0153-2027 · Sep 8, 2026 · Open, ClassifiedPERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H
- Z-1516-2025 · Mar 7, 2025 · Open, ClassifiedETHIBOND EXCEL GRN 36IN(90CM) USP2-0(M3) D/A MH
- Z-1097-2025 · Dec 20, 2024 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas
- Z-2486-2020 · Jun 10, 2020 · Open, ClassifiedPERMA-HAND/MERSILK SILK BLU and VIRGIN SILK BLU, Product Codes EH7457G EH7983G FV768 FV769 FV7771 U7058 U7059 W1782 W1784 W1819 W1820 W818 W819 W870
- Z-1841-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL Polyester Suture-ETHIBOND Green 75cm USP1 Single Armed CT-1 Product Code: X425H Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue app
- Z-1842-2019 · Apr 19, 2019 · TerminatedETHIBOND EXCEL ETHIBOND Polvester Suture-Green 75cm USP1 Single Armed CTX Product Code: X865W Product Usage: ETHIBOND EXCEL Suture is intended for use in general soft tissue appr
- Z-0747-2019 · Dec 18, 2018 · TerminatedPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approxima
- Z-1839-2015 · May 8, 2015 · TerminatedEthicon Fast Absorbing Plain Gut - MultiPass Needles, Sterile, Absorbable Surgical Suture, 5-0 (1.5 metric), 18" (45 cm), Conventional Cutting, Lot #JD6163, Exp: July 2019. Fast ab
- Z-0492-2015 · Oct 17, 2014 · TerminatedEthicon PERMA-HAND Silk siliconized black braided 10 strands per packet Ethicon LLC- Made in USA.
Common questions about the ETHIBOND EXCEL X557H
Has the ETHIBOND EXCEL X557H been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705031057562), based on the FDA recall file dated Nov 27, 2026. However, 10 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the ETHIBOND EXCEL X557H?
The FDA's MAUDE database lists 187 adverse event reports for this device, received between Apr 4, 2018 and Nov 14, 2025. In those reports, the most common specific device problems are “Material Separation” (175), “Break” (11) and “Material Frayed” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ETHIBOND EXCEL X557H?
None of the reports are classified as deaths, 0 as injuries and 187 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ETHIBOND EXCEL X557H FDA approved?
It was cleared through 510(k) K946173 on Jan 9, 1995. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the ETHIBOND EXCEL X557H MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the ETHIBOND EXCEL X557H, and is it still sold?
The FDA's device database lists Ethicon, LLC as the company responsible for it (brand ETHIBOND EXCEL, model X557H). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE devices
- FAST-FIX 72202468Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE1 death6,566 reports
- FAST-FIX 72201491Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE862 reports
- FAST-FIX 72202469Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE561 reports
- FAST-FIX 72205676Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE428 reports
- Arthrex® AR-4501Arthrex, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE400 reports
- FAST-FIX 72205324Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE381 reports
- LUPINE ORTHOCORD 210712Medos International Sàrl · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE352 reports
- FAST-FIX 72201494Smith & Nephew, Inc. · SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE249 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page