Skip to content

Ingenia recallZ-3237-2026 · Philips North America Llc · initiated Aug 21, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

FDA root cause category: Software design

What the company is doing

On August 24, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: " To prevent this issue from occurring, o Use the standard 3D Vane scan protocol from the Philips Examcard Database, or o For custom 3D Vane scan protocols, use a maximum of 480 for the reconstruction matrix and use a maximum of 167 for the number of slices. " You may continue to use affected systems in accordance with their intended use. " Circulate this notice to all users or new owners of this device so that they are aware of the issues and associated hazard/harm. " Please retain this Urgent Medical Device letter with your system(s) until the software upgrade is installed; ensure the notice is in a place likely to be seen/viewed. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the issue (reference 2026-PD-MR-006).

Recall details

Recall number
Z-3237-2026
Status
Open, Classified
Initiated by the company
Aug 21, 2026
Posted by the FDA
Sep 18, 2026
Recalling firm
Philips North America Llc
Quantity
922
Product code
LNH
FDA submissions
K183063, K193215, K213583, K251808
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia;

Product description

Ingenia Ambition S and Ingenia Ambition X; Model Numbers (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 781356 (X), 5. 782109 (X), 6. 782138 (X);

Code information (models, lots, serial numbers)

Show the full code information (7k characters)

1. Model Numbers (REF): 781359(S); UDI-DI: 884838090057; Serial Numbers: 48053, 48126, 48136, 48144, 48020, 48030, 48155, 48191, 48216, 48218, 48245, 48300, 48183, 48186, 48187, 48222, 48225, 48234, 48240, 48241, 48249, 48269, 48278, 48284, 48288, 48293, 48294, 48299, 48304, 48313, 48317, 48326, 48331, 48051, 48223, 48207, 48028, 48031, 48049, 48050, 48131, 48134, 48242, 48254, 48219, 48000, 48004, 48019, 48032, 48132, 48135, 48145, 48160, 48224, 48244, 48040, 48122, 48039, 48022, 48152, 48012, 48014, 48036, 48037, 48044, 48127, 48151, 48158, 48174, 48184, 48046, 48047, 48055, 48128, 48129, 48153, 48190, 48192, 48194, 48226, 48275, 48181, 48198, 48199, 48200, 48033, 48045, 48001, 48038, 48185, 48238, 48239, 48276, 48010, 48170, 48035, 48172, 48027, 48248, 48146, 48147, 48148, 48163, 48164, 48165, 48169, 48182, 48188, 48271, 48290, 48291, 48297, 48302, 48303, 48308, 48309, 48310, 48321, 48325, 48329, 48337, 48018, 48283, 48177, 48017, 48056, 48250, 48272, 48253, 48268, 48026, 48013, 48052, 48112, 48120, 48130, 48162, 48171, 48179, 48214, 48217, 48233, 48237, 48243, 48252, 48273, 48274, 48286, 48287, 48314, 48322, 48133, 48316; 2. Model Numbers (REF): 782108(S); UDI-DI: 884838098343; Serial Numbers: 48572, 48627, 48581, 48687, 48592, 48652, 48800, 48953, 48544, 48506, 48520, 48521, 48534, 48551, 48781, 48834, 48842, 48843, 48848, 48892, 48712, 48714, 48715, 48810, 48813, 48833, 48883, 48885, 48943, 48947, 48538, 48539, 48556, 48602, 48640, 48808, 48613, 48650, 48502, 48507, 48514, 48540, 48577, 48579, 48580, 48590, 48591, 48612, 48614, 48617, 48620, 48628, 48632, 48638, 48692, 48697, 48698, 48721, 48765, 48818, 48819, 48824, 48503, 48557, 48570, 48595, 48606, 48651, 48706, 48707, 48725, 48729, 48744, 48756, 48773, 48796, 48804, 48815, 48861, 48876, 48898, 48695, 48660, 48734, 48877, 48878, 48879, 48679, 48889, 48529, 48609, 48622, 48658, 48659, 48769, 48774, 48837, 48857, 48864, 48574, 48900, 48541, 48549, 48710, 48865, 48504, 48508, 48513, 48524, 48536, 48537, 48555, 48560, 48571, 48573, 48585, 48593, 48594, 48596, 48597, 48600, 48611, 48624, 48639, 48642, 48644, 48666, 48711, 48762, 48768, 48775, 48778, 48891, 48950, 48522, 48561, 48668, 48686, 48737, 48846, 48852, 48887, 48607, 48519, 48680, 48699, 48738, 48820, 48845, 48860, 48867, 48868, 48869, 48870, 48901, 48942, 48948, 48951, 48958, 48962, 48532, 48533, 48505, 48567, 48671, 48672, 48676, 48801, 48604, 48851, 48515, 48584, 48886, 48674, 48753, 48754, 48782, 48783, 48784, 48785, 48789, 48790, 48791, 48793, 48794, 48807, 48875, 48902, 48945, 48941, 48961, 48550, 48647, 48850, 48610, 48615, 48654, 48728, 48685, 48817, 48500, 48635, 48523, 48535, 48546, 48552, 48562, 48578, 48589, 48645, 48646, 48648, 48649, 48655, 48656, 48665, 48669, 48675, 48700, 48708, 48719, 48748, 48749, 48826, 48827, 48828, 48832, 48855, 48872, 48874, 48888, 48907, 48973, 48661, 48662, 48663, 48724, 48693, 48884, 48525, 48526, 48978; 3. Model Numbers (REF): 782139(S); UDI-DI: 884838108639; Serial Numbers: 29021, 29038, 29000, 29008, 29010, 29011, 29012, 29014, 29016, 29022, 29039, 29040, 29045, 29052, 29072, 29096, 29097, 29105, 29108, 29127, 29130, 29131, 29081, 29090, 29095, 29151, 29054, 29069, 29078, 29115, 29053, 29118, 29139, 29140, 29003, 29015, 29020, 29028, 29033, 29143, 29169, 29013, 29138, 29144, 29062, 29076, 29079, 29126, 29132, 29137, 29153, 29155, 29157, 29158, 29164, 29136, 29150, 29160, 29068, 29070, 29165, 29120, 29057, 29047, 29049, 29133, 29134, 29042, 29043, 29055, 29065, 29122; 4. Model Numbers (REF): 781356(X); UDI-DI: 884838090040; Serial Numbers: 47194, 47040, 47091, 47345, 47361, 47050, 47209, 47254, 47275, 47049, 47357, 47013, 47014, 47015, 47053, 47065, 47082, 47000, 47002, 47017, 47024, 47035, 47047, 47056, 47058, 47074, 47078, 47089, 47192, 47225, 47337, 47226, 47227, 47341, 47307, 47364, 47365, 47372, 47284, 47350, 47057, 47075, 47088, 47174, 47175, 47193, 47052, 47085, 47262, 47328, 47342, 47384, 47064, 47244, 47219, 47220, 47298, 47247, 47329, 47096, 47207, 47283, 47190, 47264, 47034, 47202, 47315, 47332, 47367, 47385, 47016, 47269, 47382, 47352, 47021, 47001, 47032, 47041, 47248, 47255, 47311, 47272, 47306, 47077, 47009, 47080, 47083, 47242, 47245, 47302, 47310, 47312, 47006, 47007, 47011, 47018, 47022, 47025, 47028, 47033, 47044, 47045, 47054, 47060, 47061, 47062, 47066, 47071, 47084, 47090, 47093, 47095, 47176, 47183, 47185, 47196, 47198, 47203, 47204, 47205, 47208, 47210, 47216, 47251, 47280, 47291, 47316, 47323, 47324, 47325, 47330, 47331, 47333, 47335, 47336, 47338, 47339, 47340, 47346, 47370, 47377, 47378, 47380, 47387; 5. Model Numbers (REF): 782109(X); UDI-DI: 884838098350; Serial Numbers: 47531, 47540, 47877, 47878, 47879, 47869, 47519, 47526, 47555, 47584, 47839, 47565, 47644, 47742, 47585, 47582, 47755, 47903, 47928, 47935, 47940, 47713, 47946, 47947, 47944, 47507, 47508, 47610, 47900, 47901, 47527, 47529, 47545, 47640, 47725, 47778, 47855, 47858, 47870, 47532, 47533, 47547, 47548, 47591, 47831, 47843, 47882, 47607, 47583, 47625, 47626, 47627, 47628, 47629, 47630, 47675, 47676, 47786, 47850, 47854, 47868, 47921, 47933, 47509, 47525, 47765, 47902, 47908, 47919, 47504, 47579, 47683, 47688, 47737, 47784, 47808, 47809, 47810, 47822, 47826, 47837, 47846, 47847, 47887, 47905, 47909, 47910, 47936, 47938, 47939, 47511, 47536, 47543, 47561, 47562, 47564, 47646, 47671, 47681, 47694, 47698, 47750, 47603, 47705, 47578, 47819, 47516, 47841, 47907, 47954, 47955, 47956, 47957, 47958, 47959, 47937, 47518, 47716, 47754, 47782, 47783, 47689, 47853, 47952, 47953, 47656, 47867, 47912, 47914, 47798, 47948, 47665, 47557, 47834, 47550, 47513, 47602, 47528, 47768, 47770, 47773, 47804, 47860, 47883, 47889, 47785, 47791, 47934, 47635, 47788, 47906, 47923, 47696, 47747, 47806, 47865, 47535, 47614, 47617, 47651, 47652, 47660, 47661, 47662, 47663, 47718, 47720, 47721, 47727, 47728, 47729, 47730, 47731, 47779, 47851, 47560, 47546, 47563, 47697, 47797, 47811, 47845, 47904, 47566, 47597, 47679, 47695, 47753, 47577, 47701, 47738, 47514, 47632, 47792, 47538, 47554, 47574, 47575, 47576, 47621, 47642, 47659, 47672, 47739, 47740, 47759, 47763, 47771, 47794, 47799, 47802, 47813, 47814, 47823, 47835, 47859, 47861, 47888, 47891, 47896, 47929, 47931, 47950; 6. Model Numbers (REF): 782138(X); UDI-DI: 884838108622; Serial Numbers: 49004, 49049, 49106, 49136, 49063, 49075, 49183, 49197, 49246, 49269, 49272, 49020, 49021, 49022, 49027, 49029, 49030, 49031, 49033, 49034, 49036, 49037, 49038, 49039, 49043, 49055, 49060, 49068, 49069, 49140, 49189, 49190, 49191, 49192, 49194, 49195, 49196, 49202, 49203, 49219, 49221, 49223, 49230, 49251, 49254, 49255, 49045, 49058, 49071, 49130, 49153, 49158, 49180, 49181, 49270, 49145, 49161, 49168, 49156, 49228, 49047, 49098, 49099, 49147, 49148, 49154, 49220, 49222, 49232, 49268, 49206, 49290, 49008, 49009, 49092, 49093, 49094, 49146, 49072, 49080, 49087, 49104, 49105, 49109, 49173, 49210, 49231, 49233;

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 99769).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page