Ingenia Ingenia Ambition S (REF 782108)Recalls & reported problems

Philips Medical Systems Nederland B.V. · Enables trained physicians to obtain cross-sectional and spectroscopic images of internal structures of the head, body or extremities, in any orientation, representing the spatial distribution of protons or other nuclei with spin. · Class 2

All models of Ingenia (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

8 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
8

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00884838098343), received between Dec 9, 2025 and Dec 9, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Device Handling Problem” (1 report).

8 FDA recall records name this device's identifier, 8 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 28, 2026Class IIOpen, Classified

    When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

    Z-3310-2026 · Root cause: Software design · Recall details · FDA record ↗

  2. Aug 21, 2026Class IIOpen, Classified

    A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

    Z-3237-2026 · Root cause: Software design · Recall details · FDA record ↗

  3. Apr 14, 2026Class IIOpen, Classified

    The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction voxel size settings in the default MRE scan protocol displaying too small.

    Z-1960-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  4. Dec 3, 2025Class IIOpen, Classified

    The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

    Z-1209-2026 · Root cause: Software design · Recall details · FDA record ↗

  5. Jun 30, 2025Class IIOpen, Classified

    The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire.

    Z-2208-2025 · Root cause: Device Design · Recall details · FDA record ↗

  6. Sep 3, 2024Class IIOpen, Classified

    Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.

    Z-3197-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  7. May 3, 2024Class IIOpen, Classified

    The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

    Z-1850-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  8. Mar 12, 2024Class IIOpen, Classified

    Patient support table floor plate may be incorrectly installed.

    Z-1568-2024 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Handling Problem1

Patient problems

  • Bone Fracture(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 9, 2025Device Handling Problem; Device Handling Problem

    PHILIPS RECEIVED A REPORT THAT THE PATIENT WAS LYING ON HIS BACK, HEAD FIRST AND HIS HAND WAS ON THE SIDE OF THE BED AND THE PATIENT WAS SCHEDULED TO UNDERGO AN IRM PROSTATE SCAN. THE EXAM WAS COMPLETED AS SCHEDULED. THE REPORTED INJURY OCCURRED AT THE BEGINNING OF THE EXAM, WHILE IN PROCESS OF TRANSFERRING ONTO THE MRI. THE PATIENT'S FINGER INJURY WAS REPORTED TO HAVE OCCURRED ON THE BED'S RAIL. AS PER THE REPORT PROVIDED, THE PATIENT GOT HIS RING FINGER STUCK IN THE TABLE RAIL AS IT MOVED TO ENTER THE TUNNEL AT THE BEGINNING OF THE EXAMINATION. THE PATIENT STATUS IS DOCUMENTED AS DISTAL PHALANX FRACTURE OF THE RING FINGER ON THE LEFT HAND. THE PHOTO DEPICTS A SIMULATION OF A HAND GRIPPING THE SIDE OF THE TABLE, WITH THE INDEX FINGER GRIPPING THE SIDE. THE REPORTED INJURY MEETS THE CRITERIA FOR SERIOUS INJURY.

    Patient problems: Bone Fracture(s)

    MAUDE report 23754292 ↗

Other System, nuclear magnetic resonance imaging devices

All System, nuclear magnetic resonance imaging devices →