DxC AU recallZ-0084-2027 · Beckman Coulter Inc. · initiated Sep 2, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
FDA root cause category: Software design
What the company is doing
On September 2, 2026, the firm began notifying customers via letters titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to inform their laboratory or medical director to assess laboratory and patient impact. For analyzers on software versions earlier than 1.5: - Users should verify that their instrument displays the intended units in the Test Menu page within the System Configuration page prior to reporting patient results. - Recall letter should be posted on or near affected analyzers if they have the affected software version. This issue is resolved in software version 1.5. If your analyzer is currently operating on an earlier software version, your Beckman Coulter service representative will contact you to schedule the upgrade to software version 1.5.
Recall details
- Recall number
- Z-0084-2027
- Status
- Open, Classified
- Initiated by the company
- Sep 2, 2026
- Posted by the FDA
- Oct 2, 2026
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 146 US, 170 OUS
- Product code
- JJE
- FDA submissions
- K220977
- Distribution
- Worldwide - US Nationwide distribution.
Product description
DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520
Code information (models, lots, serial numbers)
Software version 1.4.5 or earlier. 1. UDI-DI 14987666545058 2. UDI-DI 14987666545065
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- DxC AU C63520Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use8 recalls17 reports
- DxC AU C63519Beckman Coulter, Inc. · Analyzer, Chemistry (Photometric, Discrete), For Clinical Use6 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 99800).
- Z-0083-2027 · Open, ClassifiedDxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU Module, DxC 500i, Part No. C63521 3. Access 2 Module, DxC 500i, Par…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page