Skip to content

DxC AURecalls & reported problems across all models

Beckman Coulter, Inc. · 4 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

8 open recalls name at least one DxC AU model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
632
Deaths reported
1
Injuries reported
66
Recalls
8

We found 4 models of DxC AU by Beckman Coulter, Inc. in the FDA's device registry (GUDID). Together they are named in 632 adverse event reports in the FDA's MAUDE database, received between Jul 18, 2018 and Sep 30, 2026 — classified as 1 death, 66 injuries, 565 malfunctions.

The most frequently reported specific device problem is “Low Test Results” (302 reports).

8 FDA recall records name at least one of these models by its device identifier, 8 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 6 reportsDec 2023: 8 reportsJan 2024: 7 reportsFeb 2024: 4 reportsMar 2024: 9 reportsApr 2024: 6 reportsMay 2024: 9 reportsJun 2024: 10 reportsJul 2024: 4 reportsAug 2024: 15 reportsSep 2024: 12 reportsOct 2024: 14 reportsNov 2024: 17 reportsDec 2024: 4 reportsJan 2025: 17 reportsFeb 2025: 10 reportsMar 2025: 12 reportsApr 2025: 12 reportsMay 2025: 13 reportsJun 2025: 11 reportsJul 2025: 22 reportsAug 2025: 18 reportsSep 2025: 3 reportsOct 2025: 7 reportsNov 2025: 23 reportsDec 2025: 17 reportsJan 2026: 10 reportsFeb 2026: 14 reportsMar 2026: 17 reportsApr 2026: 16 reportsMay 2026: 17 reportsJun 2026: 11 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202623
Last 36 months. The most recent month may be incomplete.

DxC AU recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Sep 2, 2026Open, Classified

    Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and [email protected] in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.

    Z-0084-2027 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  2. Jul 15, 2026Open, Classified

    a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.

    Z-2974-2026 · Root cause: Software design (manufacturing process) · names 2 of these models · Recall details · FDA record ↗

  3. Mar 20, 2025Open, Classified

    Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.

    Z-1784-2025 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  4. Mar 17, 2025Open, Classified

    The reason for the recall is incorrect sample barcode identification and processing errors that result in the reporting of erroneous and delayed results.

    Z-1672-2025 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  5. Feb 26, 2025Open, Classified

    Beckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent tests of a calculated result are rerun, either manually ordered by the operator at the analyzer or automatically triggered by predefined rules in System Configuration, the calculated result will not be recalculated using the constituent tests rerun results. The system will report only the calculated result using the initial constituent tests results to the operator and LIS/Remisol. The issue was identified by Beckman Coulter internal engineering testing. The defect may cause a delay in reporting patient results.

    Z-1517-2025 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

  6. Jan 10, 2025Open, Classified

    Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.

    Z-1233-2025 · Root cause: Software change control · names 2 of these models · Recall details · FDA record ↗

  7. Jul 10, 2024Open, Classified

    Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.

    Z-2691-2024 · Root cause: Software design · Recall details · FDA record ↗

  8. Jun 5, 2024Open, Classified

    There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient order and a calibration order. When a patient sample and a non-barcoded calibrator tube is processed on the rack, the calibration result will be reported as a patient result. Issue 2: calibration with expired calibrator. They both may cause erroneous patient results, but the probability of serious adverse health consequences or death is unlikely.

    Z-2455-2024 · Root cause: Software design · Recall details · FDA record ↗

Most reported DxC AU problems

Device problems

  • Low Test Results302
  • High Test Results159
  • Incorrect, Inadequate or Imprecise Result or Readings139
  • Adverse Event Without Identified Device or Use Problem17
  • Imprecise or Erratic Results13
  • Low Readings3
  • Unable to Obtain Readings3
  • High Readings2

Patient problems

  • No Clinical Signs, Symptoms or Conditions541
  • No Consequences Or Impact To Patient42
  • Insufficient Information19
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available14
  • Hyponatremia14
  • No Code Available2
  • Back pain1
  • Needle Stick/Puncture1

DxC AU models

4 models, most reported first. Each model has its own page with its reports, recalls and device details.