INTERVENTIONAL ANGIOGRAPHY SYSTEM recallZ-0108-2027 · Canon Medical System, USA, INC. · initiated Aug 26, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Canon Medical System, USA, INC. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error

FDA root cause category: Process control

What the company is doing

On August 26, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction notification to affected consignees via mail. Canon ask consignees to take the following actions: 1. Should this problem occur before inspection of the tightening of the bolt(s) is implemented on your system, please stop using the device and call your service representative. 2. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. 3. A representative of Canon Medical Systems USA will contact you to schedule a time to inspect the tightening and properly fixing of the bolt(s) to correct this issue with your system(s). 4. If you have any questions regarding this matter, please contact your service representative contact our InTouch team at (800) 521-1968 or by email at [email protected]. 5. Canon respectfully requests that the attached form be completed and returned via email to [email protected].

Recall details

Recall number
Z-0108-2027
Status
Open, Classified
Initiated by the company
Aug 26, 2026
Posted by the FDA
Oct 7, 2026
Recalling firm
Canon Medical System, USA, INC.
Quantity
1 unit
Product code
OWB
FDA submissions
K210900
Distribution
US Nationwide distribution in the state of ID.

Product description

Alphenix INFX-8000C/B5

Code information (models, lots, serial numbers)

Serial no. B5F2532011/UDI: 4987670100192

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates