MiniMed Flex recallZ-0088-2027 · Medtronic MiniMed, Inc. · initiated Aug 20, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic MiniMed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.

FDA root cause category: Under Investigation by firm

What the company is doing

On 08/20/2026, Technical Support called affected customers to inform that, MiniMed identified an issue in which certain MiniMed Flex Insulin Pumps (MMT-8062) were released to customers with incomplete manufacturing testing. As a result of incomplete testing, your pump may have the potential for lack of protection from moisture exposure, slow charging, disconnection with sensor or MiniMed App, reduced ability to detect occlusion, no audio or low audio volume for alarms/alerts, or insulin under-delivery. If your pump experiences these issues, it could potentially lead to delayed therapy or risk of hypoglycemia, hyperglycemia, diabetic ketoacidosis (DKA) or death. For customers that were not able to be contacted via telephone call, an "URGENT MEDICAL DEVICE RECALL" Letter will be sent viaUSPS First-Class Mail. Customers are instructed to: 1. Stop using the Flex pump in their possession and revert to their back-up plan for insulin therapy. 2. Contact MiniMed immediately at (844) 885-0259 so arrangement can be made to ship customers a new Flex pump overnight. Upon calling, please say Flex when prompted to connect with an agent. 3. Please acknowledge that you ve read and understood this notification and have followed the actions listed by completing and returning the confirmation form, scanning the QR code below with your phone or tablet, or by visiting [https://www.minimed.com/en-us/safety-notices/notice33-letter] to acknowledge communication. For Questions or assistance - call the MiniMed 24-Hour Technical Support line at 1-800-646-4633, option 1.

Recall details

Recall number
Z-0088-2027
Status
Open, Classified
Initiated by the company
Aug 20, 2026
Posted by the FDA
Oct 5, 2026
Recalling firm
Medtronic MiniMed, Inc.
Quantity
25
Product code
QFG
FDA submissions
K253743
Distribution
U.S. Nationwide distribution in the states of AR, AZ, CO, FL, KS, KY, LA, MI, MN, NJ, NY, OH, PA, SD, TN, TX, and WV.

Product description

Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Product Description: The MiniMed Flex pump is an alternate controller enabled (ACE) pump intended for the subcutaneous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin. It can reliably and securely communicate with compatible digitally connected devices, including an integrated continuous glucose monitor (iCGM), interoperable Medtronic continuous glucose monitor (CGM), and interoperable automated glycemic controller (iAGC). The pump is intended to be used in conjunction with compatible, digitally connected medical devices for the purpose of drug delivery. The primary user interface and primary display for the MiniMed Flex pump is the MiniMed app, also the user interface for the compatible MiniMed iAGCs, which runs on a compatible iOS or Android device.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

MMT-8072 Lot Code: Serial numbers: FB00015839, FB00016511, FB00014455, FB00011553, FB00015700, FB00011357, FB00019307, FB00014766, FB00014822, FB00015828, FB00018580, FB00010465, FB00014472, FB00015793, FB00015794, FB00015796, FB00015799, FB00015801, FB00015977, FB00017271. UDI-DI/GTIN: 00763000848156 Note: Initially 25 pumps were identified as put into service, however 2 were shipped to Marketing Team Members and recovered. 3 Pumps were returned through complaint process. Final total of pumps available in the field is 20. MMT-8062 Lot Code: Serial numbers: FB00011770, FB00015162, FB00016429, FB00014760, FB00015797 UDI-DI/GTIN: 00763000848149 (Serial Numbers: FB00017271, FB00011770, FB00015162, FB00016429, FB00014760 were returned via complaint process or shipped to Marketing Team Members).

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines