MiniMed Flex MMT-8062Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-8062 FLEX US REPL · Class 2

All models of MiniMed Flex (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Aug 20, 2026Class IIOpen, Classified

    Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests. The unverified/failed pumps released to users potentially have one or more of the following issues: inoperable pump at setup, lack of protection for moisture exposure, slow battery charging, disconnection issues with other system components (i.e., continuous glucose monitoring sensor or MiniMed mobile app), reduced ability to detect occlusion events, compromised audio for pump alarms/alerts, and/or inaccurate insulin delivery.

    Z-0088-2027 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K253743 on Mar 12, 2026 (decision: Substantially Equivalent). Submitted by Medtronic Minimed, Inc. as “MiniMed Flex pump”. The FDA review took 108 days. FDA record ↗

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the MiniMed Flex MMT-8062

Has the MiniMed Flex MMT-8062 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0088-2027 (Class II), began on Aug 20, 2026: Some MiniMed Flex insulin pumps were released with missing (i.e., devices with no result or abort test results are presumed to have failed) or failed verification test results in one or more of 9 functional tests.… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the MiniMed Flex MMT-8062 FDA approved?

It was cleared through 510(k) K253743 on Mar 12, 2026. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the MiniMed Flex MMT-8062 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the MiniMed Flex MMT-8062, and is it still sold?

The FDA's device database lists Medtronic Minimed, Inc. as the company responsible for it (brand MiniMed Flex, model MMT-8062). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Alternate controller enabled insulin infusion pump devices

All Alternate controller enabled insulin infusion pump devices →